Monte Carlo source model for photon beam radiotherapy: photon source characteristics

Fix, MK; Keall, PJ; Dawson, K; Siebers, JV (2004). Monte Carlo source model for photon beam radiotherapy: photon source characteristics. Medical physics, 31(11), pp. 3106-21. College Park, Md.: American Association of Physicists in Medicine AAPM 10.1118/1.1803431

Full text not available from this repository. (Request a copy)

A major barrier to widespread clinical implementation of Monte Carlo dose calculation is the difficulty in characterizing the radiation source within a generalized source model. This work aims to develop a generalized three-component source model (target, primary collimator, flattening filter) for 6- and 18-MV photon beams that match full phase-space data (PSD). Subsource by subsource comparison of dose distributions, using either source PSD or the source model as input, allows accurate source characterization and has the potential to ease the commissioning procedure, since it is possible to obtain information about which subsource needs to be tuned. This source model is unique in that, compared to previous source models, it retains additional correlations among PS variables, which improves accuracy at nonstandard source-to-surface distances (SSDs). In our study, three-dimensional (3D) dose calculations were performed for SSDs ranging from 50 to 200 cm and for field sizes from 1 x 1 to 30 x 30 cm2 as well as a 10 x 10 cm2 field 5 cm off axis in each direction. The 3D dose distributions, using either full PSD or the source model as input, were compared in terms of dose-difference and distance-to-agreement. With this model, over 99% of the voxels agreed within +/-1% or 1 mm for the target, within 2% or 2 mm for the primary collimator, and within +/-2.5% or 2 mm for the flattening filter in all cases studied. For the dose distributions, 99% of the dose voxels agreed within 1% or 1 mm when the combined source model-including a charged particle source and the full PSD as input-was used. The accurate and general characterization of each photon source and knowledge of the subsource dose distributions should facilitate source model commissioning procedures by allowing scaling the histogram distributions representing the subsources to be tuned.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology > Medical Radiation Physics

UniBE Contributor:

Fix, Michael

ISSN:

0094-2405

ISBN:

15587664

Publisher:

American Association of Physicists in Medicine AAPM

Language:

English

Submitter:

Factscience Import

Date Deposited:

04 Oct 2013 14:56

Last Modified:

05 Dec 2022 14:17

Publisher DOI:

10.1118/1.1803431

PubMed ID:

15587664

Web of Science ID:

000225372300021

URI:

https://boris.unibe.ch/id/eprint/24114 (FactScience: 47008)

Actions (login required)

Edit item Edit item
Provide Feedback