Alleviation of chronic venous leg ulcers with a hand-held dielectric barrier discharge plasma generator (PlasmaDerm(®) VU-2010): results of a monocentric, two-armed, open, prospective, randomized and controlled trial (NCT01415622).

Brehmer, F; Haenssle, H A; Daeschlein, G; Ahmed, R; Pfeiffer, S; Görlitz, A; Simon, Dagmar; Schön, M P; Wandke, D; Emmert, S (2014). Alleviation of chronic venous leg ulcers with a hand-held dielectric barrier discharge plasma generator (PlasmaDerm(®) VU-2010): results of a monocentric, two-armed, open, prospective, randomized and controlled trial (NCT01415622). Journal of the European Academy of Dermatology and Venereology, 29(1), pp. 148-155. Blackwell 10.1111/jdv.12490

[img] Text
2014_simon_jeadv.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (545kB) | Request a copy

BACKGROUND

Cold atmospheric plasma (CAP, i.e. ionized air) is an innovating promising tool in reducing bacteria.

OBJECTIVE

We conducted the first clinical trial with the novel PlasmaDerm(®) VU-2010 device to assess safety and, as secondary endpoints, efficacy and applicability of 45 s/cm(2) cold atmospheric plasma as add-on therapy against chronic venous leg ulcers.

METHODS

From April 2011 to April 2012, 14 patients were randomized to receive standardized modern wound care (n = 7) or plasma in addition to standard care (n = 7) 3× per week for 8 weeks. The ulcer size was determined weekly (Visitrak(®) , photodocumentation). Bacterial load (bacterial swabs, contact agar plates) and pain during and between treatments (visual analogue scales) were assessed. Patients and doctors rated the applicability of plasma (questionnaires).

RESULTS

The plasma treatment was safe with 2 SAEs and 77 AEs approximately equally distributed among both groups (P = 0.77 and P = 1.0, Fisher's exact test). Two AEs probably related to plasma. Plasma treatment resulted in a significant reduction in lesional bacterial load (P = 0.04, Wilcoxon signed-rank test). A more than 50% ulcer size reduction was noted in 5/7 and 4/7 patients in the standard and plasma groups, respectively, and a greater size reduction occurred in the plasma group (plasma -5.3 cm(2) , standard: -3.4 cm(2) ) (non-significant, P = 0.42, log-rank test). The only ulcer that closed after 7 weeks received plasma. Patients in the plasma group quoted less pain compared to the control group. The plasma applicability was not rated inferior to standard wound care (P = 0.94, Wilcoxon-Mann-Whitney test). Physicians would recommend (P = 0.06, Wilcoxon-Mann-Whitney test) or repeat (P = 0.08, Wilcoxon-Mann-Whitney test) plasma treatment by trend.

CONCLUSION

Cold atmospheric plasma displays favourable antibacterial effects. We demonstrated that plasma treatment with the PlasmaDerm(®) VU-2010 device is safe and effective in patients with chronic venous leg ulcers. Thus, larger controlled trials and the development of devices with larger application surfaces are warranted.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Dermatology, Urology, Rheumatology, Nephrology, Osteoporosis (DURN) > Clinic of Dermatology

UniBE Contributor:

Simon, Dagmar

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0926-9959

Publisher:

Blackwell

Language:

English

Submitter:

Monika Schenk

Date Deposited:

16 Apr 2015 16:24

Last Modified:

05 Dec 2022 14:38

Publisher DOI:

10.1111/jdv.12490

PubMed ID:

24666170

BORIS DOI:

10.7892/boris.60528

URI:

https://boris.unibe.ch/id/eprint/60528

Actions (login required)

Edit item Edit item
Provide Feedback