Comparison of theoretical fixation stability of three devices employed in medial opening wedge high tibial osteotomy: a finite element analysis.

Golovakhа, Maxim L; Orljanski, Weniamin; Benedetto, Karl-Peter; Panchenko, Sergey; Büchler, Philippe; Henle, Philipp; Aghayev, Emin (2014). Comparison of theoretical fixation stability of three devices employed in medial opening wedge high tibial osteotomy: a finite element analysis. BMC musculoskeletal disorders, 15(230), p. 230. BioMed Central 10.1186/1471-2474-15-230

[img]
Preview
Text
Golovakhа et al. - BMC musculoskeletal disorders - 2014.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

Medial open wedge high tibial osteotomy is a well-established procedure for the treatment of unicompartmental osteoarthritis and symptomatic varus malalignment. We hypothesized that different fixation devices generate different fixation stability profiles for the various wedge sizes in a finite element (FE) analysis.

METHODS

Four types of fixation were compared: 1) first and 2) second generation Puddu plates, and 3) TomoFix plate with and 4) without bone graft. Cortical and cancellous bone was modelled and five different opening wedge sizes were studied for each model. Outcome measures included: 1) stresses in bone, 2) relative displacement of the proximal and distal tibial fragments, 3) stresses in the plates, 4) stresses on the upper and lower screw surfaces in the screw channels.

RESULTS

The highest load for all fixation types occurred in the plate axis. For the vast majority of the wedge sizes and fixation types the shear stress (von Mises stress) was dominating in the bone independent of fixation type. The relative displacements of the tibial fragments were low (in μm range). With an increasing wedge size this displacement tended to increase for both Puddu plates and the TomoFix plate with bone graft. For the TomoFix plate without bone graft a rather opposite trend was observed.For all fixation types the occurring stresses at the screw-bone contact areas pulled at the screws and exceeded the allowable threshold of 1.2 MPa for at least one screw surface. Of the six screw surfaces that were studied, the TomoFix plate with bone graft showed a stress excess of one out of twelve and without bone graft, five out of twelve. With the Puddu plates, an excess stress occurred in the majority of screw surfaces.

CONCLUSIONS

The different fixation devices generate different fixation stability profiles for different opening wedge sizes. Based on the computational simulations, none of the studied osteosynthesis fixation types warranted an intransigent full weight bearing per se. The highest fixation stability was observed for the TomoFix plates and the lowest for the first generation Puddu plate. These findings were revealed in theoretical models and need to be validated in controlled clinical settings.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Pre-clinic Human Medicine > Institute for Surgical Technology & Biomechanics ISTB [discontinued]
04 Faculty of Medicine > Pre-clinic Human Medicine > Institute for Evaluative Research into Orthopaedic Surgery

UniBE Contributor:

Büchler, Philippe, Aghayev, Emin

Subjects:

500 Science > 570 Life sciences; biology
600 Technology > 610 Medicine & health

ISSN:

1471-2474

Publisher:

BioMed Central

Language:

English

Submitter:

Philippe Büchler

Date Deposited:

16 Dec 2014 12:33

Last Modified:

05 Dec 2022 14:38

Publisher DOI:

10.1186/1471-2474-15-230

PubMed ID:

25012591

BORIS DOI:

10.7892/boris.60905

URI:

https://boris.unibe.ch/id/eprint/60905

Actions (login required)

Edit item Edit item
Provide Feedback