Systematic analysis of the intravoxel incoherent motion threshold separating perfusion and diffusion effects: Proposal of a standardized algorithm

Wurnig, Moritz C; Donati, Olivio F; Ulbrich, Erika; Filli, Lukas; Kenkel, David; Thöny, Harriet C.; Boss, Andreas (2014). Systematic analysis of the intravoxel incoherent motion threshold separating perfusion and diffusion effects: Proposal of a standardized algorithm. Magnetic resonance in medicine, 74(5), pp. 1414-1422. Wiley-Liss 10.1002/mrm.25506

[img] Text
mrm25506.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (1MB) | Request a copy

PURPOSE

To systematically evaluate the dependence of intravoxel-incoherent-motion (IVIM) parameters on the b-value threshold separating the perfusion and diffusion compartment, and to implement and test an algorithm for the standardized computation of this threshold.

METHODS

Diffusion weighted images of the upper abdomen were acquired at 3 Tesla in eleven healthy male volunteers with 10 different b-values and in two healthy male volunteers with 16 different b-values. Region-of-interest IVIM analysis was applied to the abdominal organs and skeletal muscle with a systematic increase of the b-value threshold for computing pseudodiffusion D*, perfusion fraction Fp , diffusion coefficient D, and the sum of squared residuals to the bi-exponential IVIM-fit.

RESULTS

IVIM parameters strongly depended on the choice of the b-value threshold. The proposed algorithm successfully provided optimal b-value thresholds with the smallest residuals for all evaluated organs [s/mm2]: e.g., right liver lobe 20, spleen 20, right renal cortex 150, skeletal muscle 150. Mean D* [10(-3) mm(2) /s], Fp [%], and D [10(-3) mm(2) /s] values (±standard deviation) were: right liver lobe, 88.7 ± 42.5, 22.6 ± 7.4, 0.73 ± 0.12; right renal cortex: 11.5 ± 1.8, 18.3 ± 2.9, 1.68 ± 0.05; spleen: 41.9 ± 57.9, 8.2 ± 3.4, 0.69 ± 0.07; skeletal muscle: 21.7 ± 19.0; 7.4 ± 3.0; 1.36 ± 0.04.

CONCLUSION

IVIM parameters strongly depend upon the choice of the b-value threshold used for computation. The proposed algorithm may be used as a robust approach for IVIM analysis without organ-specific adaptation. Magn Reson Med, 2014. © 2014 Wiley Periodicals, Inc.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology > DCR Magnetic Resonance Spectroscopy and Methodology (AMSM)
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > Forschungsbereich Pavillon 52 > Abt. Magnetresonanz-Spektroskopie und Methodologie, AMSM
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology

UniBE Contributor:

Thöny, Harriet C., Boss, Andreas

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0740-3194

Publisher:

Wiley-Liss

Language:

English

Submitter:

Aisha Stefania Mzinga

Date Deposited:

07 Apr 2015 10:28

Last Modified:

05 Dec 2022 14:44

Publisher DOI:

10.1002/mrm.25506

PubMed ID:

25360990

Uncontrolled Keywords:

intravoxel incoherent motion, diffusion-weighted imaging, magnetic resonance, pseudodiffusion

BORIS DOI:

10.7892/boris.66117

URI:

https://boris.unibe.ch/id/eprint/66117

Actions (login required)

Edit item Edit item
Provide Feedback