Clinical and haemodynamic outcomes in 658 patients receiving the Perceval sutureless aortic valve: early results from a prospective European multicentre study (the Cavalier Trial)†.

Laborde, François; Fischlein, Theodor; Hakim-Meibodi, Kavous; Misfeld, Martin; Carrel, Thierry; Zembala, Marian; Madonna, Francesco; Meuris, Bart; Haverich, Axel; Shrestha, Malakh (2015). Clinical and haemodynamic outcomes in 658 patients receiving the Perceval sutureless aortic valve: early results from a prospective European multicentre study (the Cavalier Trial)†. European journal of cardio-thoracic surgery, 49(3), pp. 978-986. Oxford University Press 10.1093/ejcts/ezv257

[img]
Preview
Text
ejcts.ezv257.full.pdf - Published Version
Available under License Publisher holds Copyright.

Download (258kB) | Preview

OBJECTIVES

The aim of the Cavalier trial was to evaluate the safety and performance of the Perceval sutureless aortic valve in patients undergoing aortic valve replacement (AVR). We report the 30-day clinical and haemodynamic outcomes from the largest study cohort with a sutureless valve.

METHODS

From February 2010 to September 2013, 658 consecutive patients (mean age 77.8 years; 64.4% females; mean logistic EuroSCORE 10.2%) underwent AVR in 25 European Centres. Isolated AVRs were performed in 451 (68.5%) patients with a less invasive approach in 219 (33.3%) cases. Of the total, 40.0% were octogenarians. Congenital bicuspid aortic valve was considered an exclusion criterion.

RESULTS

Implantation was successful in 628 patients (95.4%). In isolated AVR through sternotomy, the mean cross-clamp time and the cardiopulmonary bypass (CPB) time were 32.6 and 53.7 min, and with the less invasive approach 38.8 and 64.5 min, respectively. The 30-day overall and valve-related mortality rates were 3.7 and 0.5%, respectively. Valve explants, stroke and endocarditis occurred in 0.6, 2.1 and in 0.1% of cases, respectively. Preoperative mean and peak pressure gradients decreased from 44.8 and 73.24 mmHg to 10.24 and 19.27 mmHg at discharge, respectively. The mean effective orifice area improved from 0.72 to 1.46 cm(2).

CONCLUSIONS

The current 30-day results show that the Perceval valve is safe (favourable haemodynamic effect and low complication rate), and can be implanted with a fast and reproducible technique after a short learning period. Short cross-clamp and CPB times were achieved in both isolated and combined procedures. The Perceval valve represents a promising alternative to biological AVR, especially with a less invasive approach and in older patients.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Heart Surgery

UniBE Contributor:

Carrel, Thierry

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1010-7940

Publisher:

Oxford University Press

Language:

English

Submitter:

Daniela Huber

Date Deposited:

15 Jan 2016 08:23

Last Modified:

27 Feb 2024 14:28

Publisher DOI:

10.1093/ejcts/ezv257

PubMed ID:

26245628

Uncontrolled Keywords:

Aortic valve replacement; Multicentre study; Sutureless aortic valve prosthesis

BORIS DOI:

10.7892/boris.74964

URI:

https://boris.unibe.ch/id/eprint/74964

Actions (login required)

Edit item Edit item
Provide Feedback