Fattori, Giovanni; Safai, Sairos; Carmona, Pablo Fernández; Peroni, Marta; Perrin, Rosalind; Weber, Damien Charles; Lomax, Antony John (2017). Monitoring of breathing motion in image-guided PBS proton therapy: comparative analysis of optical and electromagnetic technologies. Radiation oncology, 12(1), p. 63. BioMed Central 10.1186/s13014-017-0797-9
|
Text
s13014-017-0797-9.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (1MB) | Preview |
BACKGROUND
Motion monitoring is essential when treating non-static tumours with pencil beam scanned protons. 4D medical imaging typically relies on the detected body surface displacement, considered as a surrogate of the patient's anatomical changes, a concept similarly applied by most motion mitigation techniques. In this study, we investigate benefits and pitfalls of optical and electromagnetic tracking, key technologies for non-invasive surface motion monitoring, in the specific environment of image-guided, gantry-based proton therapy.
METHODS
Polaris SPECTRA optical tracking system and the Aurora V3 electromagnetic tracking system from Northern Digital Inc. (NDI, Waterloo, CA) have been compared both technically, by measuring tracking errors and system latencies under laboratory conditions, and clinically, by assessing their practicalities and sensitivities when used with imaging devices and PBS treatment gantries. Additionally, we investigated the impact of using different surrogate signals, from different systems, on the reconstructed 4D CT images.
RESULTS
Even though in controlled laboratory conditions both technologies allow for the localization of static fiducials with sub-millimetre jitter and low latency (31.6 ± 1 msec worst case), significant dynamic and environmental distortions limit the potential of the electromagnetic approach in a clinical setting. The measurement error in case of close proximity to a CT scanner is up to 10.5 mm and precludes its use for the monitoring of respiratory motion during 4DCT acquisitions. Similarly, the motion of the treatment gantry distorts up to 22 mm the tracking result.
CONCLUSIONS
Despite the line of sight requirement, the optical solution offers the best potential, being the most robust against environmental factors and providing the highest spatial accuracy. The significant difference in the temporal location of the reconstructed phase points is used to speculate on the need to apply the same monitoring system for imaging and treatment to ensure the consistency of detected phases.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology |
UniBE Contributor: |
Weber, Damien Charles |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1748-717X |
Publisher: |
BioMed Central |
Language: |
English |
Submitter: |
Beatrice Scheidegger |
Date Deposited: |
23 Feb 2018 13:57 |
Last Modified: |
05 Dec 2022 15:07 |
Publisher DOI: |
10.1186/s13014-017-0797-9 |
PubMed ID: |
28359341 |
Uncontrolled Keywords: |
CT imaging Electromagnetic tracking Gantry Optical tracking Proton therapy Respiratory motion |