Basic concepts in metal work failure after metastatic spine tumour surgery.

Kumar, Naresh; Patel, Ravish; Wadhwa, Anshuja Charvi; Kumar, Aravind; Milavec, Helena Maria; Sonawane, Dhiraj; Singh, Gurpal; Benneker, Lorin Michael (2018). Basic concepts in metal work failure after metastatic spine tumour surgery. European spine journal, 27(4), pp. 806-814. Springer 10.1007/s00586-017-5405-z

[img]
Preview
Text
benneker2.pdf - Published Version
Available under License Publisher holds Copyright.

Download (1MB) | Preview

PURPOSE

The development of spinal implants marks a watershed in the evolution of metastatic spine tumour surgery (MSTS), which has evolved from standalone decompressive laminectomy to instrumented stabilization and decompression with reconstruction when necessary. Fusion may not be feasible after MSTS due to poor quality of graft host bed along with adjunct chemotherapy and/or radiotherapy postoperatively. With an increase in the survival of patients with spinal tumours, there is a probability of an increase in the rate of implant failure. This review aims to help establish a clear understanding of implants/constructs used in MSTS and to highlight the fundamental biomechanics of implant/construct failures.

METHODS

Published literature on implant failure after spine surgery and MSTS has been reviewed. The evolution of spinal implants and their role in MSTS has been briefly described. The review defines implant/construct failures using radiological parameters that are practical, feasible, and derived from historical descriptions. We have discussed common modes of implant/construct failure after MSTS to allow further understanding, interception, and prevention of catastrophic failure.

RESULTS

Implant failure rates in MSTS are in the range of 2-8%. Variability in patterns of failure has been observed based on anatomical region and the type of constructs used. Patients with construct/implant failures may or may not be symptomatic and present either as early (< 3months) or late failures (> 3months). It has been noted that not all the implant failures after MSTS result in revisions.

CONCLUSION

Based on the observed radiological criteria and clinical presentations, we have proposed a clinico-radiological classification for implant/construct failure after MSTS.

Item Type:

Journal Article (Review Article)

Division/Institute:

04 Faculty of Medicine > Department of Orthopaedic, Plastic and Hand Surgery (DOPH) > Clinic of Orthopaedic Surgery

UniBE Contributor:

Milavec, Helena Maria, Benneker, Lorin Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0940-6719

Publisher:

Springer

Language:

English

Submitter:

Lilianna Bolliger

Date Deposited:

25 Jan 2018 11:34

Last Modified:

05 Dec 2022 15:09

Publisher DOI:

10.1007/s00586-017-5405-z

PubMed ID:

29204734

Uncontrolled Keywords:

Asymptomatic implant failure Early failure Late failure Metastatic spine tumour surgery Symptomatic implant failure

BORIS DOI:

10.7892/boris.109973

URI:

https://boris.unibe.ch/id/eprint/109973

Actions (login required)

Edit item Edit item
Provide Feedback