Computed tomography imaging for the characterisation of drugs with radiation density measurements and HU spectroscopy.

Sieron, Dominik A; Steib, Moritz; Suter, Dominik; Obmann, Verena Carola; Huber, Adrian Thomas; Ebner, Lukas; Inderbitzin, Daniel; Christe, Andreas (2018). Computed tomography imaging for the characterisation of drugs with radiation density measurements and HU spectroscopy. Swiss medical weekly, 148, w14585. EMH Schweizerischer Ärzteverlag 10.4414/smw.2018.14585

[img]
Preview
Text
29376556.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (996kB) | Preview

AIMS OF THE STUDY

To investigate the computed tomography (CT) density of frequently administered medications (1) for the better characterisation of substances on abdominal CT, (2) to allow radiologists to narrow down possibilities in the identification of hyperdense material in the bowel and (3) to provide forensic doctors with a tool to identify gastric contents before an autopsy.

MATERIAL AND METHODS

From the list of the local hospital pharmacy, the 50 most frequently used medications were identified and scanned twice with a 128 row CT scanner (Acquillion, Toshiba, Tokyo, Japan). The protocol comprised two tube voltages of 100 kVp and 120 kVp, with a tube current of 100 mAs, a collimation of 0.5 mm and a slice thickness of 0.5 mm. Two readers were asked to measure the density (in Hounsfield units) and the noise (standard deviation of the Hounsfield units) of each pill in the two scans (100/120 kVp). After 4 weeks, both readers repeated the measurements to test repeatability (intra-rater agreement). The behaviour of each pill in hydrochloric acid (pH 2) was examined and the dissolution time was determined.

RESULTS

The most dense pill was Cordarone (7265 HU), and the least was Perenterol (529 HU), with an attenuation that was lower than fat density (<120 HU). The standard deviation of pixel density (noise) reflects inhomogeneity of the pharmacological product, varying from 9 to 1592 HU among the different pills (at 120 kVp). The absolute average HU increase per pill when changing to lower voltage was 78 ± 253 HU, with a linear fitting line with a slope of 0.21 as a constant variable in the density spectroscopy. After 4 hours in hydrochloric acid, only six tablets were still intact, including Flagyl and Dafalgan. The intra- and inter-rater agreements for all measurements were nearly perfect, with a correlation coefficient r of ≥0.99 (p <0.0001).

CONCLUSION

Our data suggest that measuring the attenuation of drugs on CT images, including the homogeneity, and applying CT spectroscopy can narrow down possible identities of the most frequently medications. Other clinicians and forensic pathologists can perform this easy measurement, as the intra- and inter-reader variability is very small.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology

UniBE Contributor:

Obmann, Verena Carola, Huber, Adrian Thomas, Ebner, Lukas, Inderbitzin, Daniel, Christe, Andreas

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1424-7860

Publisher:

EMH Schweizerischer Ärzteverlag

Language:

English

Submitter:

Nicole Rösch

Date Deposited:

18 Apr 2018 09:09

Last Modified:

05 Jun 2023 16:34

Publisher DOI:

10.4414/smw.2018.14585

PubMed ID:

29376556

BORIS DOI:

10.7892/boris.113293

URI:

https://boris.unibe.ch/id/eprint/113293

Actions (login required)

Edit item Edit item
Provide Feedback