Comparison of smartphone-based and standard base-apex electrocardiography in healthy dairy cows.

Bonelli, Francesca; Vezzosi, Tommaso; Meylan, Mireille; Nocera, Irene; Ferrulli, Vincenzo; Buralli, Carlotta; Meucci, Valentina; Tognetti, Rosalba (2019). Comparison of smartphone-based and standard base-apex electrocardiography in healthy dairy cows. Journal of veterinary internal medicine, 33(2), pp. 981-986. Wiley-Blackwell 10.1111/jvim.15396

[img]
Preview
Text
Bonelli_et_al-2019-Journal_of_Veterinary_Internal_Medicine.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial (CC-BY-NC).

Download (1MB) | Preview

BACKGROUND

There is good diagnostic accuracy of smartphone-based ECG in the evaluation of heart rate (HR), heart rhythm, and ECG values for humans, horses, dogs, and cats.

OBJECTIVES

Assess feasibility and reliability of a smartphone ECG device to evaluate HR expressed as beats per minute and ECG values in cows.

ANIMALS

Fifty-five healthy Holstein Friesian cows.

METHODS

Prospective observational study. A standard base-apex ECG was acquired for 60 seconds in each cow. A smartphone ECG tracing was recorded simultaneously using a single-lead bipolar ECG recorder. All ECGs were reviewed by 1 blinded investigator. The following ECG variables were assessed: HR, P wave duration, PR interval, QRS complex and QT interval (milliseconds), P wave, and QRS complex polarity. Agreement between standard and smartphone ECG was evaluated by Cohen's k test and the Bland-Altman test.

RESULTS

Smartphone ECG tracings were interpretable in 89% of the recordings. Minimal differences of no clinical relevance were found between standard and smartphone ECG tracings regarding HR and duration of ECG waves and interval. Good agreement found in the evaluation of QRS complex polarity (85%) but not for P wave polarity (k = 0.006). Artifacts were rare but of significantly higher frequency in smartphone ECG compared to standard ECG tracings (22% versus 0%; P = .005).

CONCLUSIONS AND CLINICAL IMPORTANCE

Good quality single-lead ECG tracings can be recorded by smartphone device in healthy cows. Smartphone tracings are reliable for evaluation of HR and selective ECG variables. Smartphone ECG can represent an additional tool for ECG screening in cows.

Item Type:

Journal Article (Original Article)

Division/Institute:

05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV)
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > Clinic for Ruminants

UniBE Contributor:

Bonelli, Francesca, Meylan, Mireille

Subjects:

600 Technology > 610 Medicine & health
600 Technology > 630 Agriculture

ISSN:

0891-6640

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Nathalie Viviane Zollinger

Date Deposited:

21 May 2019 13:06

Last Modified:

05 Dec 2022 15:25

Publisher DOI:

10.1111/jvim.15396

PubMed ID:

30592089

Uncontrolled Keywords:

ECG cow iPhone mobile validation

BORIS DOI:

10.7892/boris.125107

URI:

https://boris.unibe.ch/id/eprint/125107

Actions (login required)

Edit item Edit item
Provide Feedback