Monje, Alberto; Insua, Angel; Monje, Florencio; Muñoz, Fernando; Salvi, Giovanni Edoardo; Buser, Daniel Albin; Chappuis, Vivianne (2018). Diagnostic accuracy of the implant stability quotient in monitoring progressive peri-implant bone loss: An experimental study in dogs. Clinical oral implants research, 29(10), pp. 1016-1024. Wiley-Blackwell 10.1111/clr.13368

[img] Text
Monje_et_al-2018-Clinical_Oral_Implants_Research.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (1MB) | Request a copy

OBJECTIVES

To investigate the impact of progressive bone loss in an experimental peri-implantitis model in the dog upon the implant stability quotient (ISQ) measured in the course of induced and spontaneous conditions of disease, and to evaluate the association between the clinical parameters and ISQ.

MATERIALS AND METHODS

Seventy-two implants were placed in 12 Beagle dogs. Of these, 36 implants in six dogs were assessed during ligature-induced peri-implantitis (three timepoints) and at one timepoint following a period of spontaneous progression. The ISQ was recorded using resonance frequency analysis (RFA). Furthermore, the clinical peri-implant parameters were registered at four sites per implant at each timepoint. Marginal bone loss (MBL) was determined using computed tomography at four sites per implant and bone-to-implant contact (BIC) was assessed from histological samples. A linear regression model was estimated by generalized estimation equations (GEEs) in order to study the MBL-ISQ values at each measurement timepoint. Pearson's correlation test was applied.

RESULTS

None of the implants failed during the study period. At implant level, a strong negative correlation was found for all timepoints between ISQ and MBL (r = -0.58; p < 0.001). Accordingly, as follow-up progressed, lower ISQ and higher MBL values were observed. A prediction of MBL depending on the ISQ values and timepoints showed a decrease in one ISQ unit to be related to ~1 mm of MBL. Likewise, a statistically significant correlation was found between BIC and ISQ evaluated after spontaneous chronification of peri-implantitis (r = 0.34; p = 0.04). Nevertheless, the ISQ values failed to correlate to any of the clinical parameters recorded.

CONCLUSION

Resonance frequency analysis seems accurate in diagnosing progressive bone loss, as a statistically significant decrease in ISQ was recorded in the course of peri-implant disease. Nevertheless, the clinical relevance of this observation as a diagnostic tool is debatable, since implant stability remains high.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Research
04 Faculty of Medicine > School of Dental Medicine > Department of Oral Surgery and Stomatology
04 Faculty of Medicine > School of Dental Medicine > Department of Periodontology

UniBE Contributor:

Monje, Alberto, Salvi, Giovanni Edoardo, Buser, Daniel Albin, Chappuis, Vivianne

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0905-7161

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Doris Burri

Date Deposited:

19 Feb 2019 14:38

Last Modified:

05 Dec 2022 15:25

Publisher DOI:

10.1111/clr.13368

PubMed ID:

30267449

Uncontrolled Keywords:

dental implants diagnostic resonance frequency analysis implant stability peri-implant mucositis peri-implantitis

BORIS DOI:

10.7892/boris.125311

URI:

https://boris.unibe.ch/id/eprint/125311

Actions (login required)

Edit item Edit item
Provide Feedback