Deming, Liang; Taeschler, Dominik; Göpfert, Christine; Arnold, Patrik; Zurbuchen, Adrian; Sweda, Romy; Reichlin, Tobias; Tanner, Hildegard; Roten, Laurent; Haeberlin, Andreas (2019). Radiofrequency ablation lesion assessment using optical coherence tomography - a proof-of-concept study. Journal of cardiovascular electrophysiology, 30(6), pp. 934-940. Wiley-Blackwell 10.1111/jce.13917
Text
Liang_et_al-2019-Journal_of_Cardiovascular_Electrophysiology.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (1MB) |
BACKGROUND
Radiofrequency catheter ablation (RFA) is an effective treatment for atrial fibrillation. However, ablation lesions are usually only assessed functionally. The immediate effect of RFA on the tissue is not directly visualized. Optical coherence tomography (OCT) is an imaging technique that uses light to capture high-resolution images with histology-like quality. Therefore, it might be used for high-precision imaging of ablation lesions.
METHODS AND RESULTS
Radiofrequency ablation lesions (n = 25) were produced on the freshly excised left and right ventricular porcine endocardium. A Thermocool ST SF NAV ablation catheter (Biosense Webster Inc) and an EP-Shuttle ablation generator (Stockert GmbH) were used to produce ablation lesions with powers from 10 to 40 W (energies ranging from 100 Ws to 900 Ws). After ablation, the tissue was imaged with a swept source OCT system (at a wavelength of 1300 nm). Subsequently, the ablation lesions underwent the histological analysis. The ablation lesions could be visualized by OCT in all 17 samples with ablation powers ≥20 W, meanwhile, no lesion could be observed in the other eight samples with lower power (10 W). Lesion depths and lesion radiuses, as assessed by OCT, correlated well with those observed on the subsequent histological analysis (Spearman's r = 0.94, P < 0.001 and r = 0.84, P < 0.001). In addition, successful three-dimensional reconstructions of ablation lesions were performed.
CONCLUSION
OCT can provide a visual high-resolution assessment of ablation lesions.