Schriber, Martina; Yeung, Andy Wai Kan; Suter, Valerie G.A.; Buser, Daniel; Leung, Yiu Yan; Bornstein, Michael M. (2020). Cone beam computed tomography artefacts around dental implants with different materials influencing the detection of peri-implant bone defects. Clinical oral implants research, 31(7), pp. 595-606. Wiley-Blackwell 10.1111/clr.13596
Text
Schriber et al. 2020 COIR.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (1MB) |
OBJECTIVES
To investigate the diagnostic accuracy of cone beam computed tomography (CBCT) for the diagnosis of peri-implant bone defects of titanium (Ti), zirconium dioxide (ZrO2 ) or titanium-zirconium (Ti-Zr) alloy implants.
MATERIALS AND METHODS
Ti, Ti-Zr or ZrO2 implants with two diameters (3.3 mm, 4.1 mm) and one length (10 mm) were inserted in the angle of the mandible of six fresh defrosted pig jaws. Out of the 12 implants inserted, 6 served in the test group with standardized buccal peri-implant bone defects, whereas 6 served as control without bone defects. CBCTs were performed with three acquisition protocols (standard, high and low dose) using two devices. Four observers analysed CBCTs as follows: (a) presence of a peri-implant defect; (b) presence of peri-implant artefacts and impact on defect diagnosis; and (c) linear measurements of buccal peri-implant defect including height and width (in mm).
RESULTS
CBCT device, CBCT settings, implant material, implant diameter and observer background did not significantly influence diagnostic accuracy. The sensitivity and specificity values were high for defect detection. ZrO2 led to a lower than average diagnostic accuracy (0.781). The linear measurements of peri-implant defect were underestimated by <1 mm on average. The subjective impact of artefacts on defect diagnosis was significantly affected by implant material and observer background.
CONCLUSIONS
CBCT showed high diagnostic accuracy for peri-implant bone defect detection regardless of the device, imaging setting or implant material used. If CBCT is indicated to assess peri-implant bone disease, low dose protocols could be a promising imaging modality.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Oral Surgery and Stomatology |
UniBE Contributor: |
Schriber, Martina, Suter, Valérie, Buser, Daniel Albin |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0905-7161 |
Publisher: |
Wiley-Blackwell |
Language: |
English |
Submitter: |
Caroline Balz |
Date Deposited: |
02 Jul 2020 12:10 |
Last Modified: |
05 Dec 2022 15:39 |
Publisher DOI: |
10.1111/clr.13596 |
PubMed ID: |
32147872 |
Uncontrolled Keywords: |
CT imaging clinical assessment diagnosis imaging radiology |
BORIS DOI: |
10.7892/boris.144968 |
URI: |
https://boris.unibe.ch/id/eprint/144968 |