PURPOSE OF REVIEW

Critical appraisal of the available evidence on the self-expanding ACURATE neo transcatheter heart valve (THV) for the treatment of aortic valve disease.

RECENT FINDINGS

In an investigator-initiated, multicenter, randomized non-inferiority trial with broad inclusion criteria, ACURATE neo failed to meet non-inferiority compared with SAPIEN 3 with regard to a primary composite safety and efficacy endpoint at 30 days. The difference was driven by higher rates of moderate or severe paravalvular regurgitation and higher rates of acute kidney injury. In turn, registry data suggest that the safety and efficacy profile of the ACURATE neo is comparable to that of other commercially available devices. Randomized evidence indicated favorable hemodynamic results with large effective orifice areas and low residual gradients. The self-expanding ACURATE neo THV is associated with higher rates of residual aortic regurgitation compared to the balloon-expandable SAPIEN 3 THV. The supra-annular design with low residual gradients may be advantageous in patients with small anatomy and mild degree of calcification.

Item Type:

Journal Article (Review Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology

UniBE Contributor:

Okuno, Taishi, Lanz, Jonas, Pilgrim, Thomas

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1523-3782

Publisher:

Springer

Funders:

[UNSPECIFIED] Open

Language:

English

Submitter:

Nadia Biscozzo

Date Deposited:

04 Dec 2020 15:50

Last Modified:

05 Dec 2022 15:41

Publisher DOI:

10.1007/s11886-020-01364-4

PubMed ID:

32770431

Uncontrolled Keywords:

Balloon-expandable SCOPE Self-expanding Transcatheter aortic valve replacement, ACURATE neo Transcatheter heart valve

BORIS DOI:

10.7892/boris.147541

URI:

https://boris.unibe.ch/id/eprint/147541

Actions (login required)

Edit item Edit item
Provide Feedback