System performance evaluation of the cobas t 711 and cobas t 511 coagulation analyzers in routine laboratory settings.

Kitchen, Steve; de Maat, Moniek; Nagler, Michael; Jones, Robert; Lowe, Anna; Burden, John; Groneberg, Kai; Rozsnyai, Gergely (2020). System performance evaluation of the cobas t 711 and cobas t 511 coagulation analyzers in routine laboratory settings. Blood coagulation & fibrinolysis, 31(7), pp. 459-468. Wolters Kluwer Health 10.1097/MBC.0000000000000947

[img]
Preview
Text
System_performance_evaluation_of_the_cobas_t_711.7.pdf - Accepted Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (336kB) | Preview

: Utility of coagulation analyzers in real-world settings depends on characteristics that are often not studied comprehensively. This study aimed to investigate the analytical performance, system functionality, practicability, consistency and throughput of two new automated coagulation analyzers in routine laboratory practice. Real-world settings were simulated in three major European hemostasis laboratories and multiple assays were performed in anonymized plasma samples in parallel with routine clinical practice on the cobas t 711 (high-throughput) and cobas t 511 (mid-throughput) analyzers using activated partial thromboplastin time (aPTT), aPTT Lupus, aPTT Screen, Antithrombin (AT), D-Dimer, Fibrinogen, Prothrombin Time (PT)-derived Fibrinogen, PT Owren, PT Rec (recombinant human thromboplastin reagent) and Thrombin Time assays. Precision was tested in a 21-day experiment and accuracy was compared with reference methods of the same laboratory. A number of experiments simulated challenging real-life situations. Pearson's correlation coefficient was more than 0.98 in all assays. Across assays, coefficients of variation ranged from 0.0 to 1.5% for intermediate precision; 0.2 to 3.0% for repeatability and 0.4 to 3.7% for total precision. Good between-run comparability was seen when testing samples under random conditions. Calculated maximum throughput was 197 and 387-402 tests/h for the cobas t 511 and 711 analyzers, respectively. Practicability met or exceeded user expectations in 98% of cases. In a simulated real-life setting of three major laboratories, the new cobas t 511 and cobas t 711 coagulation analyzers demonstrated a good functionality, practicability and performance and the throughput was high.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Institute of Clinical Chemistry

UniBE Contributor:

Nagler, Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1473-5733

Publisher:

Wolters Kluwer Health

Language:

English

Submitter:

Karin Balmer

Date Deposited:

09 Nov 2020 12:09

Last Modified:

05 Dec 2022 15:41

Publisher DOI:

10.1097/MBC.0000000000000947

PubMed ID:

32826393

BORIS DOI:

10.7892/boris.147749

URI:

https://boris.unibe.ch/id/eprint/147749

Actions (login required)

Edit item Edit item
Provide Feedback