In vitro hemodynamic evaluation of ventricular suction conditions of the EVAHEART ventricular assist pump

Ferreira, Antonio L; Yu, Yih-Choung; Arnold, Dorian K; Vandenberghe, Stijn; Grashow, Jonathan; Kitano, Tomoya; Borzelleca, David; Antaki, James F (2012). In vitro hemodynamic evaluation of ventricular suction conditions of the EVAHEART ventricular assist pump. International journal of artificial organs, 35(4), pp. 263-271. Milano: Wichtig Editore 10.5301/ijao.5000066

Full text not available from this repository.

Purpose: Mismatches between pump output and venous return in a continuous-flow ventricular assist device may elicit episodes of ventricular suction. This research describes a series of in vitro experiments to characterize the operating conditions under which the EVAHEART centrifugal blood pump (Sun Medical Technology Research Corp., Nagano, Japan) can be operated with minimal concern regarding left ventricular (LV) suction. Methods: The pump was interposed into a pneumatically driven pulsatile mock circulatory system (MCS) in the ventricular apex to aorta configuration. Under varying conditions of preload, afterload, and systolic pressure, the speed of the pump was increased step-wise until suction was observed. Identification of suction was based on pump inlet pressure. Results: In the case of reduced LV systolic pressure, reduced preload (=10 mmHg), and afterload (=60 mmHg), suction was observed for speeds =2,200 rpm. However, suction did not occur at any speed (up to a maximum speed of 2,400 rpm) when preload was kept within 10-14 mmHg and afterload =80 mmHg. Although in vitro experiments cannot replace in vivo models, the results indicated that ventricular suction can be avoided if sufficient preload and afterload are maintained. Conclusion: Conditions of hypovolemia and/or hypotension may increase the risk of suction at the highest speeds, irrespective of the native ventricular systolic pressure. However, in vitro guidelines are not directly transferrable to the clinical situation; therefore, patient-specific evaluation is recommended, which can be aided by ultrasonography at various points in the course of support.

Item Type:

Journal Article (Original Article)

Division/Institute:

10 Strategic Research Centers > ARTORG Center for Biomedical Engineering Research > ARTORG Center - Cardiovascular Engineering (CVE)

UniBE Contributor:

Vandenberghe, Stijn

ISSN:

0391-3988

Publisher:

Wichtig Editore

Language:

English

Submitter:

Factscience Import

Date Deposited:

04 Oct 2013 14:39

Last Modified:

05 Dec 2022 14:12

Publisher DOI:

10.5301/ijao.5000066

PubMed ID:

22505201

Web of Science ID:

000305879300003

URI:

https://boris.unibe.ch/id/eprint/15824 (FactScience: 223288)

Actions (login required)

Edit item Edit item
Provide Feedback