Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the Biograph Vision Quadra with the Biograph Vision PET/CT.

Alberts, Ian; Hünermund, Jan-Niklas; Prenosil, George; Mingels, Clemens; Bohn, Karl Peter; Viscione, Marco; Sari, Hasan; Vollnberg, Bernd; Shi, Kuangyu; Afshar-Oromieh, Ali; Rominger, Axel (2021). Clinical performance of long axial field of view PET/CT: a head-to-head intra-individual comparison of the Biograph Vision Quadra with the Biograph Vision PET/CT. European journal of nuclear medicine and molecular imaging, 48(8), pp. 2395-2404. Springer 10.1007/s00259-021-05282-7

[img]
Preview
Text
259_2021_Article_5282.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

PURPOSE

To investigate the performance of the new long axial field-of-view (LAFOV) Biograph Vision Quadra PET/CT and a standard axial field-of-view (SAFOV) Biograph Vision 600 PET/CT (both: Siemens Healthineers) system using an intra-patient comparison.

METHODS

Forty-four patients undergoing routine oncological PET/CT were prospectively included and underwent a same-day dual-scanning protocol following a single administration of either 18F-FDG (n = 20), 18F-PSMA-1007 (n = 16) or 68Ga-DOTA-TOC (n = 8). Half the patients first received a clinically routine examination on the SAFOV (FOVaxial 26.3 cm) in continuous bed motion and then immediately afterwards on the LAFOV system (10-min acquisition in list mode, FOVaxial 106 cm); the second half underwent scanning in the reverse order. Comparisons between the LAFOV at different emulated scan times (by rebinning list mode data) and the SAFOV were made for target lesion integral activity, signal to noise (SNR), target lesion to background ratio (TBR) and visual image quality.

RESULTS

Equivalent target lesion integral activity to the SAFOV acquisitions (16-min duration for a 106 cm FOV) were obtained on the LAFOV in 1.63 ± 0.19 min (mean ± standard error). Equivalent SNR was obtained by 1.82 ± 1.00 min LAFOV acquisitions. No statistically significant differences (p > 0.05) in TBR were observed even for 0.5 min LAFOV examinations. Subjective image quality rated by two physicians confirmed the 10 min LAFOV to be of the highest quality, with equivalence between the LAFOV and the SAFOV at 1.8 ± 0.85 min. By analogy, if the LAFOV scans were maintained at 10 min, proportional reductions in applied radiopharmaceutical could obtain equivalent lesion integral activity for activities under 40 MBq and equivalent doses for the PET component of <1 mSv.

CONCLUSION

Improved image quality, lesion quantification and SNR resulting from higher sensitivity were demonstrated for an LAFOV system in a head-to-head comparison under clinical conditions. The LAFOV system could deliver images of comparable quality and lesion quantification in under 2 min, compared to routine SAFOV acquisition (16 min for equivalent FOV coverage). Alternatively, the LAFOV system could allow for low-dose examination protocols. Shorter LAFOV acquisitions (0.5 min), while of lower visual quality and SNR, were of adequate quality with respect to target lesion identification, suggesting that ultra-fast or low-dose acquisitions can be acceptable in selected settings.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Clinic of Nuclear Medicine

UniBE Contributor:

Alberts, Ian Leigh, Hünermund, Jan-Niklas, Prenosil, George, Mingels, Clemens, Bohn, Karl Peter, Shi, Kuangyu, Afshar Oromieh, Ali, Rominger, Axel Oliver

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1619-7089

Publisher:

Springer

Language:

English

Submitter:

Daria Vogelsang

Date Deposited:

10 Jan 2022 08:48

Last Modified:

05 Dec 2022 15:54

Publisher DOI:

10.1007/s00259-021-05282-7

PubMed ID:

33797596

Uncontrolled Keywords:

Digital PET PET/CT Positron-emission-tomography Total-body Ultra-long FOV PET Whole-body

BORIS DOI:

10.48350/161331

URI:

https://boris.unibe.ch/id/eprint/161331

Actions (login required)

Edit item Edit item
Provide Feedback