Alberts, Ian; Sachpekidis, Christos; Prenosil, George; Viscione, Marco; Bohn, Karl Peter; Mingels, Clemens; Shi, Kuangyu; Afshar-Oromieh, Ali; Rominger, Axel (2021). Digital PET/CT allows for shorter acquisition protocols or reduced radiopharmaceutical dose in [18F]-FDG PET/CT. Annals of nuclear medicine, 35(4), pp. 485-492. Springer 10.1007/s12149-021-01588-6
|
Text
12149_2021_Article_1588.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (800kB) | Preview |
Purpose: To establish the feasibility of shorter acquisition times (and by analogy, applied activity) on tumour detection and lesion contrast in digital PET/CT.
Methods: Twenty-one randomly selected patients who underwent oncological [18F]-FDG PET/CT on a digital PET/CT were retrospectively evaluated. Scan data were anonymously obtained and reconstructed in list-mode acquisition for a standard 2 min/bed position (bp), 1 min/bp and 30 s/bp (100%, 50% and 25% time or applied activity, respectively). Scans were randomized and read by two nuclear medicine physicians in a consensus read. Readers were blind to clinical details. Scans were evaluated for the number of pathological lesions detected. Measured uptake for lesions was evaluated by maximum and mean standardized uptake value (SUVmax and SUVmean, respectively) and tumour-to-backround ratio (TBR) were compared. Agreement between the three acquisitions was compared by Krippendorf's alpha.
Results: Overall n = 100 lesions were identified in the 2 min and 1 min/bp acquisitions and n = 98 lesions in the 30 s/bp acquisitions. Agreement between the three acquisitions with respect to lesion number and tumour-to-background ratio showed almost perfect agreement (K's α = 0.999). SUVmax, SUVmean and TBR likewise showed > 98% agreement, with longer acquisitions being associated with slightly higher mean TBR (2 min/bp 7.94 ± 4.41 versus 30 s/bp 7.84 ± 4.22, p < 0.05).
Conclusion: Shorter acquisition times have traditionally been associated with reduced lesion detectability or the requirement for larger amounts of radiotracer activity. These data confirm that this is not the case for new-generation digital PET scanners, where the known higher sensitivity results in clinically adequate images for shorter acquisitions. Only a small variation in the semi-quantitative parameters SUVmax, SUVmean and TBR was seen, confirming that either reduction of acquisition time or (by analogy) applied activity can be reduced as much as 75% in digital PET/CT without apparent clinical detriment.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Clinic of Nuclear Medicine |
UniBE Contributor: |
Alberts, Ian Leigh, Sachpekidis, Christos, Prenosil, George, Bohn, Karl Peter, Shi, Kuangyu, Afshar Oromieh, Ali, Rominger, Axel Oliver |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1864-6433 |
Publisher: |
Springer |
Language: |
English |
Submitter: |
Daria Vogelsang |
Date Deposited: |
13 Jan 2022 09:53 |
Last Modified: |
05 Dec 2022 15:56 |
Publisher DOI: |
10.1007/s12149-021-01588-6 |
PubMed ID: |
33550515 |
Uncontrolled Keywords: |
Digital PET; List mode acquisition; PET/CT; Positron emission tomography |
BORIS DOI: |
10.48350/162358 |
URI: |
https://boris.unibe.ch/id/eprint/162358 |