Ebel, Florian; Wanderer, Stefan; Jesse, Marvin; Schär, Ralph T.; Zubak, Irena; Ulrich, Christian T.; Raabe, Andreas (2022). A standardized model for in vitro testing of sutures and patches for watertight dural closure. Journal of neurosurgery, 136(5), pp. 1485-1494. American Association of Neurological Surgeons 10.3171/2021.5.JNS21369
Text
_19330693_-_Journal_of_Neurosurgery__A_standardized_model_for_in_vitro_testing_of_sutures_and_patches_for_watertight_dural_closure.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (8MB) |
OBJECTIVE
CSF leaks are common complications of spinal and cranial surgeries. Several dural grafts and suture techniques are available to achieve watertight dural closure, but the effectiveness of these techniques remains unclear. The authors developed a standardized in vitro model to test available grafts and suture techniques alone or in combination to find the technique with the most watertight dural closure.
METHODS
A fluid chamber with a dural fixation device, infusion pump, pressure gauge, and porcine pericardium as a dural equivalent was assembled to provide the reusable device for testing. The authors performed dural closure in 4 different fashions, as follows: A) using running versus simple interrupted suture technique and different suture materials to close a 3-cm incision; B) selecting commonly used sealants and dural patches in combination with a running suture; C) performing duraplasty (1.5 × 1.5-cm square defect) with different dural substitutes in a stand-alone fashion; and D) performing duraplasty with different dural substitutes in a double-layer fashion. Each technique was tested 6 times. The hydrostatic burst pressure (BP) was measured and compared using the Kruskal-Wallis test or the Mann-Whitney U-test. Values are reported as mean ± SD.
RESULTS
There was no significant difference between the running and simple interrupted suture technique (p = 0.79). Adding a patch or sealant to a suture resulted in a 1.7- to 14-fold higher BP compared to solitary suture closure (36.2 ± 24.27 cm H2O and 4.58 ± 1.41 cm H2O, respectively; p < 0.001). The highest BP was achieved by adding DuraSeal or TachoSil (82.33 ± 12.72 cm H2O and 74.17 ± 12.64 cm H2O, respectively). For closing a square defect, using a double-layer duraplasty significantly increased BP by a factor of 4-12 compared to a single-layer duraplasty (31.71 ± 12.62 cm H2O vs 4.19 ± 0.88 cm H2O, respectively; p < 0.001). The highest BP was achieved with the combination of Lyomesh and TachoSil (43.67 ± 11.45 cm H2O).
CONCLUSIONS
A standardized in vitro model helps to objectify the watertightness of dural closure. It allows testing of sutures and dural grafts alone or in combination. In the authors' testing, a running 6-0 monofilament polypropylene suture combined with DuraSeal or TachoSil was the technique achieving the highest BP. For the duraplasty of square defects, the double-layer technique showed the highest efficacy.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Head Organs and Neurology (DKNS) > Clinic of Neurosurgery |
UniBE Contributor: |
Ebel, Florian Marc, Jesse, Christopher Marvin, Schär, Ralph Thomas, Zubak, Irena, Ulrich, Christian Thomas (A), Raabe, Andreas |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0022-3085 |
Publisher: |
American Association of Neurological Surgeons |
Language: |
English |
Submitter: |
Nicole Söll |
Date Deposited: |
13 Jan 2022 08:49 |
Last Modified: |
18 Jun 2023 00:11 |
Publisher DOI: |
10.3171/2021.5.JNS21369 |
PubMed ID: |
34624865 |
Uncontrolled Keywords: |
cerebrospinal fluid leakage dural sealant duraplasty primary dural closure surgical technique |
BORIS DOI: |
10.48350/162706 |
URI: |
https://boris.unibe.ch/id/eprint/162706 |