High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry.

Kovacs, Boldizsar; Burri, Haran; Buehler, Andres; Reek, Sven; Sticherling, Christian; Schaer, Beat; Linka, Andre; Ammann, Peter; Müller, Andreas S; Dzemali, Omer; Kobza, Richard; Schindler, Matthias; Haegeli, Laurent; Mayer, Kurt; Eriksson, Urs; Herrera-Siklody, Claudia; Reichlin, Tobias; Steffel, Jan; Saguner, Ardan M and Duru, Firat (2021). High Incidence of Inappropriate Alarms in Patients with Wearable Cardioverter-Defibrillators: Findings from the Swiss WCD Registry. Journal of clinical medicine, 10(17) MDPI 10.3390/jcm10173811

[img]
Preview
Text
High_Incidence_of_Inappropriate_Alarms_in_Patients_with_Wearable_Cardioverter-Defibrillators_Findings_from_the_Swiss_WCD_Registry.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

The wearable cardioverter defibrillator (WCD) uses surface electrodes to detect arrhythmia before initiating a treatment sequence. However, it is also prone to inappropriate detection due to artefacts.

OBJECTIVE

The aim of this study is to assess the alarm burden in patients and its impact on clinical outcomes.

METHODS

Patients from the nationwide Swiss WCD Registry were included. Clinical characteristics and data were obtained from the WCDs. Arrhythmia recordings ≥30 s in length were analysed and categorized as VT/VF, atrial fibrillation (AF), supraventricular tachycardia (SVT) or artefact.

RESULTS

A total of 10653 device alarms were documented in 324 of 456 patients (71.1%) over a mean WCD wear-time of 2.0 ± 1.6 months. Episode duration was 30 s or more in 2996 alarms (28.2%). One hundred and eleven (3.7%) were VT/VF episodes. The remaining recordings were inappropriate detections (2736 (91%) due to artefacts; 117 (3.7%) AF; 48 (1.6%) SVT). Two-hundred and seven patients (45%) had three or more alarms per month. Obesity was significantly associated with three or more alarms per month (p = 0.01, 27.7% vs. 15.9%). High alarm burden was not associated with a lower average daily wear time (20.8 h vs. 20.7 h, p = 0.785) or a decreased implantable cardioverter defibrillator implantation rate after stopping WCD use (48% vs. 47.3%, p = 0.156).

CONCLUSIONS

In patients using WCDs, alarms emitted by the device and impending inappropriate shocks were frequent and most commonly caused by artefacts. A high alarm burden was associated with obesity but did not lead to a decreased adherence.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology

UniBE Contributor:

Reichlin, Tobias Roman

Subjects:

600 Technology > 610 Medicine & health

ISSN:

2077-0383

Publisher:

MDPI

Language:

English

Submitter:

Vjollca Coli

Date Deposited:

17 Jan 2022 15:26

Last Modified:

05 Dec 2022 16:00

Publisher DOI:

10.3390/jcm10173811

PubMed ID:

34501258

Uncontrolled Keywords:

alarm inappropriate therapy obesity outcome wearable cardioverter-defibrillator

BORIS DOI:

10.48350/163326

URI:

https://boris.unibe.ch/id/eprint/163326

Actions (login required)

Edit item Edit item
Provide Feedback