A realistic validation study of a new nitrogen multiple-breath washout system

Singer, Florian; Houltz, Birgitta; Latzin, Philipp; Robinson, Paul; Gustafsson, Per (2012). A realistic validation study of a new nitrogen multiple-breath washout system. PLoS ONE, 7(4), e36083. Lawrence, Kans.: Public Library of Science 10.1371/journal.pone.0036083

[img]
Preview
Text
journal.pone.0036083.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (272kB) | Preview

Background

For reliable assessment of ventilation inhomogeneity, multiple-breath washout (MBW) systems should be realistically validated. We describe a new lung model for in vitro validation under physiological conditions and the assessment of a new nitrogen (N2)MBW system.
Methods

The N2MBW setup indirectly measures the N2 fraction (FN2) from main-stream carbon dioxide (CO2) and side-stream oxygen (O2) signals: FN2 = 1−FO2−FCO2−FArgon. For in vitro N2MBW, a double chamber plastic lung model was filled with water, heated to 37°C, and ventilated at various lung volumes, respiratory rates, and FCO2. In vivo N2MBW was undertaken in triplets on two occasions in 30 healthy adults. Primary N2MBW outcome was functional residual capacity (FRC). We assessed in vitro error (√[difference]2) between measured and model FRC (100–4174 mL), and error between tests of in vivo FRC, lung clearance index (LCI), and normalized phase III slope indices (Sacin and Scond).
Results

The model generated 145 FRCs under BTPS conditions and various breathing patterns. Mean (SD) error was 2.3 (1.7)%. In 500 to 4174 mL FRCs, 121 (98%) of FRCs were within 5%. In 100 to 400 mL FRCs, the error was better than 7%. In vivo FRC error between tests was 10.1 (8.2)%. LCI was the most reproducible ventilation inhomogeneity index.
Conclusion

The lung model generates lung volumes under the conditions encountered during clinical MBW testing and enables realistic validation of MBW systems. The new N2MBW system reliably measures lung volumes and delivers reproducible LCI values.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Gynaecology, Paediatrics and Endocrinology (DFKE) > Clinic of Paediatric Medicine

UniBE Contributor:

Singer, Florian, Latzin, Philipp

ISSN:

1932-6203

Publisher:

Public Library of Science

Language:

English

Submitter:

Anette van Dorland

Date Deposited:

04 Oct 2013 14:41

Last Modified:

05 Dec 2022 14:12

Publisher DOI:

10.1371/journal.pone.0036083

PubMed ID:

22558338

Web of Science ID:

000305336000123

BORIS DOI:

10.7892/boris.16544

URI:

https://boris.unibe.ch/id/eprint/16544 (FactScience: 224199)