Collision cross sections of negative cluster ions of phosphoric acid in N2 determined by drift tube ion mobility and their use in travelling wave ion mobility

Lavanant, Hélène; Groessl, Michael; Afonso, Carlos (2019). Collision cross sections of negative cluster ions of phosphoric acid in N2 determined by drift tube ion mobility and their use in travelling wave ion mobility. International journal of mass spectrometry, 442, pp. 14-22. Elsevier 10.1016/j.ijms.2019.04.009

[img] Text
1-s2.0-S1387380618304020-main.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (1MB) | Request a copy

Travelling wave ion mobility usually uses a calibration procedure to determine collision cross sections. Here, we investigate phosphoric acid cluster anions as possible calibrants to increase the pool of reference ions for singly and multiply charged anions with nitrogen as drift gas.

Phosphoric acid solutions (5–50 mM) in water and acetonitrile were analyzed using electrospray ionization in the negative ion mode. Collision cross sections were obtained using an IMS-TOF drift tube ion mobility mass spectrometer (Tofwerk, Thun, Switzerland) (DTIMS-MS) on which we varied the electric field. Twenty-six collision cross sections from DTIMS-MS in nitrogen gas, termed DTCCS(N2), were obtained for singly, doubly and triply charged phosphoric acid cluster anions which range from 146 to 530 Å2. Reduced mobilities K0 ranged from 0.767 to 1.435 cm2 V−1 s−1 for singly charged cluster anions, 1.177 to 1.006 cm2 V−1 s−1 and 1.250 to 1.0143 cm2 V−1 s−1 for doubly and triply charged cluster anions, respectively. The DTCCS(N2) were found fitted to a power function 80n0.51 where n is the aggregation number.

Ion mobility mass spectrometry measurements were also carried out on a travelling wave ion mobility mass spectrometry (TWIMS-MS) device (SYNAPT G2 HDMS, Waters Corp., Manchester, UK). The DTCCS(N2) were well correlated to the drift times measured on the TWIM instrument (R2 above 0.9999). The correlation function was found to depend on the charge state of the ions. Accuracy tests evidenced a systematic error of 9–11 Å2 (4–6%) on singly charged ions of polyalanine, dextran and hexakis(fluoroalkoxy)phosphazine compared to literature values. Although absolute values were shifted, CCS differences were preserved.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Dermatology, Urology, Rheumatology, Nephrology, Osteoporosis (DURN) > Clinic of Nephrology and Hypertension
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > Unit Childrens Hospital > Forschungsgruppe Nephrologie / Hypertonie

UniBE Contributor:

Grössl, Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1387-3806

Publisher:

Elsevier

Language:

English

Submitter:

Michael Grössl

Date Deposited:

14 Apr 2022 15:48

Last Modified:

05 Dec 2022 16:18

Publisher DOI:

10.1016/j.ijms.2019.04.009

BORIS DOI:

10.48350/169049

URI:

https://boris.unibe.ch/id/eprint/169049

Actions (login required)

Edit item Edit item
Provide Feedback