Soft tissue response to different abutment materials: A controlled and randomized human study using an experimental model.

Enkling, Norbert; Marder, Michael; Bayer, Stefan; Götz, Werner; Stoilov, Milan; Kraus, Dominik (2022). Soft tissue response to different abutment materials: A controlled and randomized human study using an experimental model. Clinical oral implants research, 33(6), pp. 667-679. Wiley-Blackwell 10.1111/clr.13932

[img]
Preview
Text
Clinical_Oral_Implants_Res_-_2022_-_Enkling_-_Soft_tissue_response_to_different_abutment_materials_A_controlled_and.pdf - Accepted Version
Available under License Publisher holds Copyright.

Download (13MB) | Preview

OBJECTIVES

Aim of this study was to compare the soft-tissue response to implant abutments made of titanium, zirconia, zirconia veneered with feldspar ceramics and PEEK by various clinical, histological, microbiological and molecular biological markers in an experimental model.

MATERIALS AND METHODS

40 experimental one-piece healing abutments of four different materials were mounted on bone-level implants in 20 volunteering patients (split-mouth design). After a three months period of open healing, clinical parameters at the abutments were assessed and adjacent mucosa was sampled for inflammatory cytokine mRNA-concentrations and histological analysis by a novel method. In addition, PISF-samples were obtained for the analysis of periodonto-pathogenic bacteria counts and active MMP-8 levels. Marginal bone level change was measured by intra oral radiographs.

RESULTS

Abutments of the different materials did not exhibit significant differences regarding clinical parameters, pathogenic bacteria counts or pro-inflammatory cytokine concentrations. Likewise, no significant differences were detected regarding soft tissue morphology or bone level change. Compared to titanium abutments, significantly less mononuclear inflammatory cells were detected in the mucosa at abutments made of zirconia veneered with feldspar ceramics.

CONCLUSIONS

All examined abutment materials exhibited a similar soft tissue response compared to titanium and histological data did not reveal early signs of elevated inflammation caused by PEEK- and feldspar-veneered zirconia abutments. Due to the short observation period and the small sample size, a final conclusion on the long term suitability of those abutment materials cannot be drawn. However, based on the presented data, we consider further studies on that subject as appropriate.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology

UniBE Contributor:

Enkling, Norbert

ISSN:

0905-7161

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Pubmed Import

Date Deposited:

26 Apr 2022 10:04

Last Modified:

27 Apr 2023 00:25

Publisher DOI:

10.1111/clr.13932

PubMed ID:

35467040

Uncontrolled Keywords:

PEEK abutment materials feldspar ceramics histological analysis peri-implant soft tissue soft tissue integration soft tissue response titanium zirconia

BORIS DOI:

10.48350/169512

URI:

https://boris.unibe.ch/id/eprint/169512

Actions (login required)

Edit item Edit item
Provide Feedback