Pileggi, Marco; Ventura, Elisa; Di Napoli, Alberto; Piantanida, Renato; Muto, Mario; Cardia, Andrea; Cianfoni, Alessandro (2022). Gantry-needle-target alignment technique for CT-guided needle approaches to the skull base and cranio-cervical junction. Neuroradiology, 64(10), pp. 2039-2047. Springer-Verlag 10.1007/s00234-022-03005-9
Text
Pileggi2022_Article_Gantry-needle-targetAlignmentT.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Authors hold Copyright Download (1MB) |
PURPOSE
CT-guided percutaneous procedures involving the skull base and atlanto-axial cervical spine pose particular challenges due to high density of vital vascular and nervous structures and because the ideal needle trajectory often has a cranio-caudal obliquity different from the axial scan plane. We describe how the variable CT gantry tilt, combined with gantry-needle-target alignment technique, is used to obtain precise and safe needle placement in conventional and non-conventional approaches to the skull base and the atlanto-axial spine.
METHODS
We retrospectively analyzed consecutive CT-guided needle accesses to the skull base and atlanto-axial spine performed for tissue sampling through fine-needle aspirates and core biopsies, cementoplasty of neoplastic lytic lesions of atlanto-axial spine, pain management injections, and dural puncture for cerebro-spinal fluid sampling. All the accesses were performed with the gantry-needle-target alignment technique. Procedural complications were recorded.
RESULTS
Thirty-nine CT-guided procedures were analyzed. Paramaxillary approach was used in 15 cases, postero-lateral in 11, subzygomatic in 3. Nine non-conventional approach were performed: submastoid in 3 cases, suprazygomatic in 2, trans-nasal in 2, trans-mastoid in 1, and trans-auricular in 1. Two peri-procedural complications occurred: one asymptomatic and one resolved within 24 h. All the procedures were successfully completed with successful needle access to the target.
CONCLUSION
The gantry tilt and gantry-needle-target alignment technique allows to obtain double-oblique needle accesses for CT-guided procedures involving the skull base and atlanto-axial cervical spine, minimizing uncertainty of needle trajectory and obtaining safe needle placement in conventional and non-conventional approaches.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic and Interventional Neuroradiology |
UniBE Contributor: |
Cianfoni, Alessandro |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0028-3940 |
Publisher: |
Springer-Verlag |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
06 Jul 2022 10:38 |
Last Modified: |
05 Dec 2022 16:21 |
Publisher DOI: |
10.1007/s00234-022-03005-9 |
PubMed ID: |
35788697 |
Uncontrolled Keywords: |
CT-guided interventions Skull base |
BORIS DOI: |
10.48350/171110 |
URI: |
https://boris.unibe.ch/id/eprint/171110 |