Vertical head rotation as major source of differences between time-separated digital cephalometric radiographs of patients acquired in one cephalostat X-ray device.

Schulze, Ralf K W; Linnerth, Lea K M (2022). Vertical head rotation as major source of differences between time-separated digital cephalometric radiographs of patients acquired in one cephalostat X-ray device. BMC medical imaging, 22(1), p. 208. BioMed Central 10.1186/s12880-022-00935-x

[img]
Preview
Text
s12880-022-00935-x.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

The purpose of the retrospective study was to analyze the reliability and repeatability of specific landmark-positions used in cephalometry to determine the major sources of absolute landmark position differences for repeated, time-separated (time-point I and II) digital cephalometric radiographs (CEPH) of the same patients.

METHODS

100 pairs of CPEHs from the database of a sample of adult patients (18 to 28 years) were analyzed by one calibrated observer and three landmark points (Sella: S, Nasion: N, Subspinale: A) were digitally marked using ImageJ-software. The coordinates of these points entered the evaluation using displacement vectors as primary endpoints between the coordinates of the landmarks in the two images as well as SNA-angles and the angle ω of SN relative to the floor.

RESULTS

Displacement vectors between CEPHI and CEPHII were rather large (N: 7.95 ± 4.85 mm, S: 5.34 ± 3.50 mm, A: 4.81 ± 3.95 mm. SNA was rather stable between the two sequential radiographs (mean difference: 0.002° ± 1.85°). and did not correlate with age of the patient (SNAI: spearman-Rho: 0.0239, p = 0.8134; SNAII : spearman-Rho: 0.0244, p = 0.8096). Although the vertical angle ω did not differ between CEPHI and CEPHII (mean difference: 0.4° ± 4.7°, pwilcoxon = 0.8155), it showed a quadratic relationship (pF-statistic: < 2.2e-16) with the length of the displacement vector N.

CONCLUSION

The significantly varying location of the reference points S, N and A between time-separated CEPHs of one patient can largely be explained by different angulation (head rotation within the sagittal plane) of the Frankfurt plane to the floor (horizontal plane).

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Oral Surgery and Stomatology
04 Faculty of Medicine > School of Dental Medicine

UniBE Contributor:

Schulze, Ralf Kurt Willy

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1471-2342

Publisher:

BioMed Central

Language:

English

Submitter:

Pubmed Import

Date Deposited:

28 Nov 2022 11:06

Last Modified:

05 Dec 2022 16:29

Publisher DOI:

10.1186/s12880-022-00935-x

PubMed ID:

36434538

Uncontrolled Keywords:

Cephalometric radiography Cephalometry Head position Repeated assessments

BORIS DOI:

10.48350/175185

URI:

https://boris.unibe.ch/id/eprint/175185

Actions (login required)

Edit item Edit item
Provide Feedback