Limitations of transcutaneous carbon dioxide monitoring in apneic oxygenation.

Schweizer, Thilo; Hartwich, Volker; Riva, Thomas; Kaiser, Heiko; Theiler, Lorenz; Greif, Robert; Nabecker, Sabine (2023). Limitations of transcutaneous carbon dioxide monitoring in apneic oxygenation. PLoS ONE, 18(6), e0286038. Public Library of Science 10.1371/journal.pone.0286038

[img]
Preview
Text
journal.pone.0286038.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

BACKGROUND

High-flow nasal oxygenation is increasingly used during sedation procedures and general anesthesia in apneic patients. Transcutaneous CO2 (ptcCO2)-monitoring is used to monitor hypercapnia. This study investigated ptcCO2-monitoring during apneic oxygenation.

METHODS

We included 100 patients scheduled for elective surgery under general anesthesia in this secondary analysis of a randomized controlled trial. Before surgery, we collected ptcCO2 measured by TCM4 and TCM5 monitors and arterial blood gas (ABG) measurements every two minutes during 15 minutes of apnea. Bland-Altman plots analyzed agreement between measurement slopes; linear mixed models estimated the different measuring method effect, and outlined differences in slope and offset between transcutaneous and arterial CO2 partial pressures.

RESULTS

Bland-Altman plots showed a bias in slope (95% confidence intervals) between ABG and TCM4-measurements of 0.05mmHg/min (-0.05 to 0.15), and limits of agreement were -0.88mmHg/min (-1.06 to -0.70) and 0.98mmHg/min (0.81 to 1.16). Bias between ABG and TCM5 was -0.14mmHg/min (-0.23 to -0.04), and limits of agreement were -0.98mmHg/min (-1.14 to -0.83) and 0.71mmHg/min (0.55 to 0.87). A linear mixed model (predicting the CO2-values) showed an offset between arterial and transcutaneous measurements of TCM4 (-15.2mmHg, 95%CI: -16.3 to -14.2) and TCM5 (-19.1mmHg, -20.1 to -18.0). Differences between the two transcutaneous measurements were statistically significant.

CONCLUSIONS

Substantial differences were found between the two transcutaneous measurement systems, and between them and ABG. Transcutaneous CO2 monitoring cannot replace arterial CO2-monitoring during apneic oxygenation. In clinical settings with rapidly changing CO2-values, arterial blood gas measurements are needed to reliably assess the CO2-partial pressure in blood.

TRIAL REGISTRATION

ClinicalTrials.gov (NCT03478774).

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > Clinic and Policlinic for Anaesthesiology and Pain Therapy > Partial clinic Insel
04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > Clinic and Policlinic for Anaesthesiology and Pain Therapy

UniBE Contributor:

Schweizer, Thilo, Riva, Thomas, Kaiser, Heiko Andreas, Greif, Robert, Nabecker, Sabine

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1932-6203

Publisher:

Public Library of Science

Language:

English

Submitter:

Pubmed Import

Date Deposited:

05 Jun 2023 16:24

Last Modified:

06 Sep 2024 11:54

Publisher DOI:

10.1371/journal.pone.0286038

PubMed ID:

37262066