Loebner, Hannes A; Frauchiger, Daniel; Mueller, Silvan; Guyer, Gian; Mackeprang, Paul-Henry; Stampanoni, Marco F M; Fix, Michael K; Manser, Peter; Bertholet, Jenny (2023). Technical note: Feasibility of gating for dynamic trajectory radiotherapy - Mechanical accuracy and dosimetric performance. Medical physics, 50(10), pp. 6535-6542. American Association of Physicists in Medicine AAPM 10.1002/mp.16533
|
Text
Medical_Physics_-_2023_-_Loebner_-_Technical_note_Feasibility_of_gating_for_dynamic_trajectory_radiotherapy_Mechanical.pdf - Published Version Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND). Download (1MB) | Preview |
BACKGROUND
Dynamic trajectory radiotherapy (DTRT) extends state-of-the-art volumetric modulated arc therapy (VMAT) by dynamic table and collimator rotations during beam-on. The effects of intrafraction motion during DTRT delivery are unknown, especially regarding the possible interplay between patient and machine motion with additional dynamic axes.
PURPOSE
To experimentally assess the technical feasibility and quantify the mechanical and dosimetric accuracy of respiratory gating during DTRT delivery.
METHODS
A DTRT and VMAT plan are created for a clinically motivated lung cancer case and delivered to a dosimetric motion phantom (MP) placed on the table of a TrueBeam system using Developer Mode. The MP reproduces four different 3D motion traces. Gating is triggered using an external marker block, placed on the MP. Mechanical accuracy and delivery time of the VMAT and DTRT deliveries with and without gating are extracted from the logfiles. Dosimetric performance is assessed by means of gamma evaluation (3% global/2 mm, 10% threshold).
RESULTS
The DTRT and VMAT plans are successfully delivered with and without gating for all motion traces. Mechanical accuracy is similar for all experiments with deviations <0.14° (gantry angle), <0.15° (table angle), <0.09° (collimator angle) and <0.08 mm (MLC leaf positions). For DTRT (VMAT), delivery times are 1.6-2.3 (1.6- 2.5) times longer with than without gating for all motion traces except one, where DTRT (VMAT) delivery is 5.0 (3.6) times longer due to a substantial uncorrected baseline drift affecting only DTRT delivery. Gamma passing rates with (without) gating for DTRT/VMAT were ≥96.7%/98.5% (≤88.3%/84.8%). For one VMAT arc without gating it was 99.6%.
CONCLUSION
Gating is successfully applied during DTRT delivery on a TrueBeam system for the first time. Mechanical accuracy is similar for VMAT and DTRT deliveries with and without gating. Gating substantially improved dosimetric performance for DTRT and VMAT.