A precise and simple isodose-volume-based verification method for HDR and LDR brachytherapy plans.

Schiefer, H; Heinze, S; Glatzer, M (2023). A precise and simple isodose-volume-based verification method for HDR and LDR brachytherapy plans. Brachytherapy, 22(3), pp. 400-406. Elsevier 10.1016/j.brachy.2022.12.007

[img] Text
1-s2.0-S1538472122004111-main.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (1MB)

PURPOSE

The American Association of Physicists in Medicine (AAPM) code of practice for brachytherapy physics recommends performing an independent treatment time calculation. For this we implemented an easy to use isodose-based verification method for HDR (high-dose-rate) and LDR (low-dose-rate) brachytherapy plans.

MATERIAL AND METHODS

Dose-volume-based methods have been developed for Ir-192-based high-dose-rate (HDR) and I-125 prostate low-dose-rate (LDR) brachytherapy. They allow checking the integral dwell time or activity when the volume of a suitable isodose is known. The verification method was validated for 55 clinical HDR and 243 clinical LDR plans.

RESULTS

For HDR brachytherapy, the mean absolute difference between the estimated and calculated integral dwell time was 0.8% ± 1.0% (n = 30) with a single-source path and 2.7% ± 1.1% (n = 25) for multiple source paths. The corresponding value for LDR brachytherapy was 1.8% ± 2.0% (n = 243). In HDR brachytherapy, the verification method depends slightly on the plan class when considering one or more than one source paths. Good agreement between the estimated and calculated integral dwell times was obtained based on the 2 Gy isodose. Unlike HDR brachytherapy, the parameters used in the verification method for LDR brachytherapy plan verification strongly depend on the type of seed distribution. So, we recommend using an isodose at the prescribed dose for prostate HDR therapy.

CONCLUSIONS

Isodose-based verification methods are precise, do not presuppose dedicated tools, and are simple to implement in clinical practice.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology

UniBE Contributor:

Glatzer, Markus

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1873-1449

Publisher:

Elsevier

Language:

English

Submitter:

Basak Ginsbourger

Date Deposited:

17 Nov 2023 11:40

Last Modified:

17 Nov 2023 11:49

Publisher DOI:

10.1016/j.brachy.2022.12.007

PubMed ID:

36635203

Uncontrolled Keywords:

HDR brachytherapy Isodose LDR brachytherapy Verification

BORIS DOI:

10.48350/189091

URI:

https://boris.unibe.ch/id/eprint/189091

Actions (login required)

Edit item Edit item
Provide Feedback