Two-screw osteosynthesis of the mandibular condylar head with different screw materials: a finite element analysis.

Schönegg, Daphne; Koch, Andreas; Müller, Günter T; Blumer, Michael; Wagner, Maximilian E H (2024). Two-screw osteosynthesis of the mandibular condylar head with different screw materials: a finite element analysis. Computer methods in biomechanics and biomedical engineering, 27(7), pp. 878-882. Taylor & Francis 10.1080/10255842.2023.2209247

[img]
Preview
Text
Two-screw_osteosynthesis_of_the_mandibular_condylar_head_with_different_screw_materials_a_finite_element_analysis.pdf - Published Version
Available under License Creative Commons: Attribution-Noncommercial-No Derivative Works (CC-BY-NC-ND).

Download (1MB) | Preview

This study compared the biomechanical behavior of titanium, magnesium, and polylactic acid screws for two-screw osteosynthesis of mandibular condylar head fractures using finite element analysis. Von Mises stress distribution, fracture displacement, and fragment deformation were evaluated. Titanium screws performed the best in terms of carrying the highest load, resulting in the least fracture displacement and fragment deformation. Magnesium screws showed intermediate results, while PLA screws were found to be unsuitable with stress values exceeding their tensile strength. These findings suggest that magnesium alloys could be considered a suitable alternative to titanium screws in mandibular condylar head osteosynthesis.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Head Organs and Neurology (DKNS) > Clinic of Craniomaxillofacial Surgery

UniBE Contributor:

Blumer, Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1025-5842

Publisher:

Taylor & Francis

Language:

English

Submitter:

Claudine Howald

Date Deposited:

08 Dec 2023 07:56

Last Modified:

27 Apr 2024 00:12

Publisher DOI:

10.1080/10255842.2023.2209247

PubMed ID:

37154519

Uncontrolled Keywords:

Finite element analysis magnesium mandibular condylar head screw osteosynthesis poly-lactic acid titanium

BORIS DOI:

10.48350/189107

URI:

https://boris.unibe.ch/id/eprint/189107

Actions (login required)

Edit item Edit item
Provide Feedback