PURPOSE

The physical properties of yttrium-90 (90Y) allow for imaging with positron emission tomography/computed tomography (PET/CT). The increased sensitivity of long axial field-of-view (LAFOV) PET/CT scanners possibly allows to overcome the small branching ratio for positron production from 90Y decays and to improve for the post-treatment dosimetry of 90Y of selective internal radiation therapy.

METHODS

For the challenging case of an image quality body phantom, we compare a full Monte Carlo (MC) dose calculation with the results from the two commercial software packages Simplicit90Y and Hermes. The voxel dosimetry module of Hermes relies on the 90Y images taken with a LAFOV PET/CT, while the MC and Simplicit90Y dose calculations are image independent.

RESULTS

The resulting doses from the MC calculation and Simplicit90Y agree well within the error margins. The image-based dose calculation with Hermes, however, consistently underestimates the dose. This is due to the mismatch of the activity distribution in the PET images and the size of the volume of interest. We found that only for the smallest phantom sphere there is a statistically significant dependence of the Hermes dose on the image reconstruction parameters and scan time.

CONCLUSION

Our study shows that Simplicit90Y's local deposition model can provide a reliable dose estimate. On the other hand, the image based dose calculation suffers from the suboptimal reconstruction of the 90Y distribution in small structures.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Clinic of Nuclear Medicine

UniBE Contributor:

Mercolli, Lorenzo, Zeimpekis, Konstantinos, Prenosil, George, Rathke, Hendrik Gabriel, Rominger, Axel Oliver, Shi, Kuangyu

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1724-191X

Publisher:

Elsevier

Language:

English

Submitter:

Pubmed Import

Date Deposited:

29 Jan 2024 14:51

Last Modified:

10 Feb 2024 00:16

Publisher DOI:

10.1016/j.ejmp.2024.103296

PubMed ID:

38281409

Uncontrolled Keywords:

Dosimetry Long axial field-of-view PET/CT Monte Carlo simulations Yttrium-90

BORIS DOI:

10.48350/192205

URI:

https://boris.unibe.ch/id/eprint/192205

Actions (login required)

Edit item Edit item
Provide Feedback