Reliability of a face scanner in measuring the vertical dimension of occlusion.

Srinivasan, Murali; Berisha, Florentin; Bronzino, Innocenzo; Kamnoedboon, Porawit; Leles, Claudio Rodrigues (2024). Reliability of a face scanner in measuring the vertical dimension of occlusion. (In Press). Journal of dentistry, 146(105016), p. 105016. Elsevier Science 10.1016/j.jdent.2024.105016

[img]
Preview
Text
1-s2.0-S0300571224001866-main.pdf - Accepted Version
Available under License Creative Commons: Attribution (CC-BY).

Download (1MB) | Preview

OBJECTIVE

This study evaluated the reliability of a face scanner in measuring the vertical dimension of occlusion (VDO).

METHODS

Fully dentate volunteers (n=20; mean-age=30.0±10.7 years) were recruited. Clinical facial measurements were obtained using a digital caliper and a face scanner (Class 1 LASER, Obiscanner, Fifthingenium, Italy). The scans were imported into a mesh-processing software, and the distances were measured digitally. Measurements were obtained for each participant with the jaws positioned in maximal intercuspation (MI) and with increased vertical distances of 2, 4, and 6 mm. Vertical and horizontal measures were obtained using facial anatomical landmarks: Glabella (GL), Pronasale (PrN), Subnasale (SbN), inferior border of the right and left Alare, Labiale superius (Ls), right and left Cheilion (Ch), Soft Pogonion (SPg), right and left Tragus of the ear (Tr), for all selected vertical positions. Data analysis included intra-class correlation coefficient (ICC), pairwise comparison tests, Bland-Altman plots, and Passing-Bablok regression.

RESULTS

120 VDO measurements (clinical=60, digital=60) were recorded by two independent evaluators. Mean differences between digital and clinical measurements ranged from 0.054±0.14mm to 0.203±0.13mm. All parameters were strongly correlated (r>0.93; p<0.001). ICC estimates revealed excellent reliability, and the measuring procedure yields the same results on repeated trials irrespective of the raters and measurement methods. Bland-Altman plots revealed a difference between digital and clinical measurements of 1.7% for the vertical measurements. Regression analysis revealed no significant proportional difference between the two methods, so both can be used interchangeably.

CONCLUSIONS

The findings of this study demonstrate that VDO can be measured accurately from face scans using 3D mesh-processing software and that even small changes in the VDO could be detected using the digital methods.

CLINICAL SIGNIFICANCE

Findings provide evidence about the reliability of a digital method for jaw relation registrations and may be applied towards incorporating this method into clinical workflows for CAD-CAM dentures.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine

UniBE Contributor:

Leles, Cláudio Rodrigues

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0300-5712

Publisher:

Elsevier Science

Language:

English

Submitter:

Pubmed Import

Date Deposited:

30 Apr 2024 12:55

Last Modified:

14 May 2024 00:17

Publisher DOI:

10.1016/j.jdent.2024.105016

PubMed ID:

38679136

Uncontrolled Keywords:

3D Face scan Digital complete dentures Geriatric Dentistry Jaw relations Removable prosthodontics Vertical dimension of occlusion

BORIS DOI:

10.48350/196343

URI:

https://boris.unibe.ch/id/eprint/196343

Actions (login required)

Edit item Edit item
Provide Feedback