Capek, Lukas; Huber, Adrian Thomas; Reineke, David; Dobner, Stephan; Hunziker, Lukas Christoph; Schnegg, Bruno (2024). Outflow Graft Obstruction due to Local Aortic Dissection After Implantation of Left Ventricle Assist Device (HeartMate 3). (In Press). ASAIO journal Lippincott 10.1097/MAT.0000000000002222
|
Text
outflow_graft_obstruction_due_to_local_aortic.477.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (1MB) | Preview |
Left ventricular assist devices (LVADs) improve symptoms and outcomes in patients with advanced heart failure. We report the case of a patient with a freshly implanted HeartMate 3 LVAD, suffering abruptly on postoperative day 55 from pejoration of his heart failure with multiple episodes of low-flow alarm. Outflow graft obstruction (OGO) due to local aortic dissection was diagnosed with multimodality imaging. After a multidisciplinary discussion, a surgical approach was decided, and the patient benefited from a revision of his outflow graft.
Item Type: |
Journal Article (Further Contribution) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology 04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology |
UniBE Contributor: |
Huber, Adrian Thomas, Dobner, Stephan, Hunziker Munsch, Lukas Christoph, Schnegg, Bruno |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1058-2916 |
Publisher: |
Lippincott |
Language: |
English |
Submitter: |
Pubmed Import |
Date Deposited: |
13 May 2024 15:52 |
Last Modified: |
28 May 2024 12:33 |
Publisher DOI: |
10.1097/MAT.0000000000002222 |
PubMed ID: |
38728741 |
BORIS DOI: |
10.48350/196689 |
URI: |
https://boris.unibe.ch/id/eprint/196689 |