Intrapatient reproducibility of the BISxp monitor

Niedhart, Dagmar J; Kaiser, Heiko A; Jacobsohn, Eric; Hantler, Charles B; Evers, Alex S; Avidan, Michael S (2006). Intrapatient reproducibility of the BISxp monitor. Anesthesiology, 104(2), pp. 242-8. Hagerstown, Md.: Lippincott Williams & Wilkins 10.1097/00000542-200602000-00007

Full text not available from this repository. (Request a copy)

BACKGROUND: The Bispectral Index (BIS) reportedly reflects anesthetic depth. It is recommended that anesthetic agents should be titrated to maintain the BIS between 40 and 60 arbitrary BIS units during anesthesia. For anesthesia providers to follow this recommendation, the monitor should be predictably affected by different anesthetic agents and have good interpatient and intrapatient reproducibility. The authors hypothesized that when two BISxp devices (Aspect Medical Systems, Newton, MA) are placed concurrently on the same patient, their readings are concordant throughout the anesthetic period. METHODS: Simultaneous BIS recordings from two BISxp monitors were obtained during anesthesia at 5-s intervals from 12 participants. RESULTS: In total 22,860 concurrent paired BIS readings were obtained. For 10.7% of the time, there were sustained periods of 30 s or greater where the readings suggested a different depth of anesthesia. For 6% of the time, there were sustained periods of 30 s or greater where the readings differed by 10 or more arbitrary BIS units. The regression coefficient (R) for the two devices was 0.65 (range, 0.35-0.92). There was zero bias between the devices, and the 95% limits of agreement ranged between -18 and +17. CONCLUSION: A conflicting anesthetic management was suggested by the simultaneous BIS readings 10.7% of the time. These results suggest that BISxp does not always provide a reproducible single number. Anesthesia providers should not rely exclusively on the BIS reading when assessing depth of anesthesia.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Intensive Care, Emergency Medicine and Anaesthesiology (DINA) > Clinic and Policlinic for Anaesthesiology and Pain Therapy

UniBE Contributor:

Kaiser Niedhart, Dagmar Julika

ISSN:

0003-3022

ISBN:

16436841

Publisher:

Lippincott Williams & Wilkins

Language:

English

Submitter:

Jeannie Wurz

Date Deposited:

04 Oct 2013 14:49

Last Modified:

05 Dec 2022 14:15

Publisher DOI:

10.1097/00000542-200602000-00007

PubMed ID:

16436841

Web of Science ID:

000234985500005

URI:

https://boris.unibe.ch/id/eprint/20436 (FactScience: 3733)

Actions (login required)

Edit item Edit item
Provide Feedback