Use of gold markers for setup in image-guided fractionated high-dose-rate brachytherapy as a monotherapy for prostate cancer

Ghadjar, Pirus; Gwerder, Nicole; Madlung, Axel; Behrensmeier, Frank; Thalmann, George N; Mini, Roberto; Aebersold, Daniel M (2009). Use of gold markers for setup in image-guided fractionated high-dose-rate brachytherapy as a monotherapy for prostate cancer. Strahlentherapie und Onkologie, 185(11), pp. 731-5. Heidelberg: Urban and Vogel 10.1007/s00066-009-2007-7

[img]
Preview
Text
66_2009_Article_2007.pdf - Published Version
Available under License Publisher holds Copyright.

Download (273kB) | Preview

BACKGROUND AND PURPOSE: In order to use a single implant with one treatment plan in fractionated high-dose-rate brachytherapy (HDR-B), applicator position shifts must be corrected prior to each fraction. The authors investigated the use of gold markers for X-ray-based setup and position control between the single fractions. PATIENTS AND METHODS: Caudad-cephalad movement of the applicators prior to each HDR-B fraction was determined on radiographs using two to three gold markers, which had been inserted into the prostate as intraprostatic reference, and one to two radiopaque-labeled reference applicators. 35 prostate cancer patients, treated by HDR-B as a monotherapy between 10/2003 and 06/2006 with four fractions of 9.5 Gy each, were analyzed. Toxicity was scored according to the CTCAE Score, version 3.0. Median follow-up was 3 years. RESULTS: The mean change of applicators positions compared to baseline varied substantially between HDR-B fractions, being 1.4 mm before fraction 1 (range, -4 to 2 mm), -13.1 mm before fraction 2 (range, -36 to 0 mm), -4.1 mm before fraction 3 (range, -21 to 9 mm), and -2.6 mm at fraction 4 (range, -16 to 9 mm). The original position of the applicators could be readjusted easily prior to each fraction in every patient. In 18 patients (51%), the applicators were at least once readjusted > 10 mm, however, acute or late grade > or = 2 genitourinary toxicity was not increased (p = 1.0) in these patients. CONCLUSION: Caudad position shifts up to 36 mm were observed. Gold markers represent a valuable tool to ensure setup accuracy and precise dose delivery in fractionated HDR-B monotherapy of prostate cancer.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology
04 Faculty of Medicine > Department of Dermatology, Urology, Rheumatology, Nephrology, Osteoporosis (DURN) > Clinic of Urology
04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology > Medical Radiation Physics

UniBE Contributor:

Ghadjar, Pirus, Madlung, Axel, Behrensmeier, Frank, Thalmann, George, Mini, Roberto, Aebersold, Daniel Matthias

ISSN:

0179-7158

Publisher:

Urban and Vogel

Language:

English

Submitter:

Factscience Import

Date Deposited:

04 Oct 2013 15:12

Last Modified:

02 Mar 2023 23:23

Publisher DOI:

10.1007/s00066-009-2007-7

PubMed ID:

19899006

Web of Science ID:

000271623000004

BORIS DOI:

10.7892/boris.31640

URI:

https://boris.unibe.ch/id/eprint/31640 (FactScience: 196283)

Actions (login required)

Edit item Edit item
Provide Feedback