Schindera, Sebastian T; Nelson, Rendon C; Yoshizumi, Terry; Toncheva, Greta; Nguyen, Giao; DeLong, David M; Szucs-Farkas, Zsolt (2009). Effect of automatic tube current modulation on radiation dose and image quality for low tube voltage multidetector row CT angiography: phantom study. Academic radiology, 16(8), pp. 997-1002. Philadelphia, Pa.: Elsevier 10.1016/j.acra.2009.02.021
Full text not available from this repository.RATIONALE AND OBJECTIVES: To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography. MATERIALS AND METHODS: An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR(2)/ED. RESULTS: The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P > .05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P > .05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001). CONCLUSIONS: As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology |
UniBE Contributor: |
Schindera, Sebastian Tobias, Szücs Farkas, Zsolt |
ISSN: |
1076-6332 |
Publisher: |
Elsevier |
Language: |
English |
Submitter: |
Factscience Import |
Date Deposited: |
04 Oct 2013 15:13 |
Last Modified: |
05 Dec 2022 14:22 |
Publisher DOI: |
10.1016/j.acra.2009.02.021 |
PubMed ID: |
19409820 |
Web of Science ID: |
000268298200014 |
URI: |
https://boris.unibe.ch/id/eprint/32196 (FactScience: 197177) |