Schöllhorn, Bernhard; Bürki, Alexander; Stahl, Christina; Howard, J.; Forterre, Franck (2013). Comparison of the biomechanical properties of a ventral cervical intervertebral anchored fusion device with locking plate fixation applied to cadaveric canine cervical spines. Veterinary surgery, 42(7), pp. 825-831. Wiley-Blackwell 10.1111/j.1532-950X.2013.12044.x
Text
vsu12044.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (522kB) |
OBJECTIVE: To evaluate fixation properties of a new intervertebral anchored fusion device and compare these with ventral locking plate fixation. STUDY DESIGN: In vitro biomechanical evaluation. ANIMALS: Cadaveric canine C4-C7 cervical spines (n = 9). METHODS: Cervical spines were nondestructively loaded with pure moments in a nonconstraining testing apparatus to induce flexion/extension while angular motion was measured. Range of motion (ROM) and neutral zone (NZ) were calculated for (1) intact specimens, (2) specimens after discectomy and fixation with a purpose-built intervertebral fusion cage with integrated ventral fixation, and (3) after removal of the device and fixation with a ventral locking plate. RESULTS: Both fixation techniques resulted in a decrease in ROM and NZ (P < .001) compared with the intact segments. There were no significant differences between the anchored spacer and locking plate fixation. CONCLUSION: An anchored spacer appears to provide similar biomechanical stability to that of locking plate fixation.