Impact of changes in haematocrit level and platelet count on thromboelastometry parameters

Nagler, Michael; Kathriner, Silvio; Bachmann, Lucas M.; Wuillemin, Walter A. (2013). Impact of changes in haematocrit level and platelet count on thromboelastometry parameters. Thrombosis research, 131(3), pp. 249-253. Elsevier 10.1016/j.thromres.2013.01.009

[img] Text
2013 Thromb res_ROTEM Hct PLT.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (340kB)

INTRODUCTION

To what extent haematocrit levels (Hct) and platelet counts (PLT) influence the measurement of parameters of thromboelastometry when assessed with the ROTEM® device is unclear. We investigated to what extent thromboelastometry measurements depend on Hct and PLT.

MATERIALS AND METHODS

Whole blood samples were taken for in-vitro preparations of mixtures with three different levels of PLT and a varying Hct. Maximum clot firmness (MCF), clotting time (CT), clot formation time (CFT) and alpha angle (α) for INTEM, EXTEM, FIBTEM and APTEM was recorded.

RESULTS

Measurements depended substantially on Hct and PLT. MCF readings were systematically lower with increasing Hct (0.2 vs. 0.4: -7.8 (-8.3 to -7.2); p<0.001, 0.2 vs. 0.55: -14.5 (-17.3 to -14.3); p<0.001) but higher with increasing PLT (50 vs. 125×10(9)/l: 8.2 (4.2 to 12.3); p=0.005, 50 vs. 250×10(9)/l: 12.0 (7.2 to 16.8); p=0.002). CT readings were systematically higher with increasing Hct (0.2 vs. 0.4: 9.2 (6.2 to 12.1); p=0.001, 0.2 vs. 0.55: 38.2 (21.5 to 54.9); p=0.003) while increasing PLT had no influence. CFT readings were also systematically higher with increasing Hct (0.2 vs. 0.4: 83.8 (40.2 to 127.6); p=0.006, 0.2 vs. 0.55: 226.2 (110.7 to 341.7); p=0.006) but systematically lower with increasing PLT (50 vs. 125×10(9)/l: -144.0 (-272.3 to -15.6); p=0.036, 50 vs. 250×10(9)/l: -189.2 (-330.4 to -48.0); p=0.02); readings of the alpha angle showed a similar pattern.

CONCLUSIONS

Our results suggest that readings of thromboelastometry parameters need to be adjusted by Hct and PLT to avoid potential confounding and miss-interpretations in clinical practice.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Haematology and Central Haematological Laboratory

UniBE Contributor:

Nagler, Michael

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0049-3848

Publisher:

Elsevier

Language:

English

Submitter:

Verena Zwahlen

Date Deposited:

16 Jun 2014 14:43

Last Modified:

05 Dec 2022 14:31

Publisher DOI:

10.1016/j.thromres.2013.01.009

PubMed ID:

23347682

BORIS DOI:

10.7892/boris.47408

URI:

https://boris.unibe.ch/id/eprint/47408

Actions (login required)

Edit item Edit item
Provide Feedback