Coronary artery stents: influence of adaptive statistical iterative reconstruction on image quality using 64-HDCT

Gebhard, Cathérine; Fiechter, Michael; Fuchs, Tobias A; Stehli, Julia; Müller, Ennio; Stähli, Barbara E.; Gebhard, Caroline E.; Ghadri, Jelena R.; Klaeser, Bernd; Gaemperli, Oliver; Kaufmann, Philipp A. (2013). Coronary artery stents: influence of adaptive statistical iterative reconstruction on image quality using 64-HDCT. European heart journal - cardiovascular imaging, 14(10), pp. 969-977. Oxford University Press 10.1093/ehjci/jet013

[img] Text
jet013.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.
Author holds Copyright

Download (798kB)

OBJECTIVE: The assessment of coronary stents with present-generation 64-detector row computed tomography (HDCT) scanners is limited by image noise and blooming artefacts. We evaluated the performance of adaptive statistical iterative reconstruction (ASIR) for noise reduction in coronary stent imaging with HDCT.

METHODS AND RESULTS: In 50 stents of 28 patients (mean age 64 ± 10 years) undergoing coronary CT angiography (CCTA) on an HDCT scanner the mean in-stent luminal diameter, stent length, image quality, in-stent contrast attenuation, and image noise were assessed. Studies were reconstructed using filtered back projection (FBP) and ASIR-FBP composites. ASIR resulted in reduced image noise vs. FBP (P < 0.0001). Two readers graded the CCTA stent image quality on a 4-point Likert scale and determined the proportion of interpretable stent segments. The best image quality for all clinical images was obtained with 40 and 60% ASIR with significantly larger luminal area visualization compared with FBP (+42.1 ± 5.4% with 100% ASIR vs. FBP alone; P < 0.0001) while the stent length was decreased (-4.7 ± 0.9%, <P = 0.002) and volume measurements were unaffected.

CONCLUSION: Reconstruction of CCTA from HDCT using 40 and 60% ASIR incrementally improves intra-stent luminal area, diameter visualization, and image quality compared with FBP reconstruction.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Clinic of Nuclear Medicine
04 Faculty of Medicine > Department of Head Organs and Neurology (DKNS) > Clinic of Neurosurgery

UniBE Contributor:

Fiechter, Michael, Klaeser, Bernd

Subjects:

600 Technology > 610 Medicine & health
600 Technology

ISSN:

2047-2412

Publisher:

Oxford University Press

Language:

English

Submitter:

Franziska Nicoletti

Date Deposited:

25 Jun 2014 17:10

Last Modified:

06 Dec 2022 15:30

Publisher DOI:

10.1093/ehjci/jet013

PubMed ID:

23428650

Uncontrolled Keywords:

Adaptive statistical iterative reconstruction, Coronary artery stent, Coronary CT angiography

BORIS DOI:

10.48350/48293

URI:

https://boris.unibe.ch/id/eprint/48293

Actions (login required)

Edit item Edit item
Provide Feedback