Weber, Damien Charles; Vallet, Veronique; Molineu, Andrea; Melidis, Christos; Teglas, Vanda; Naudy, Suzanne; Moeckli, Raphael; Followill, David S; Hurkmans, Coen W (2014). IMRT credentialing for prospective trials using institutional virtual phantoms: results of a joint European Organization for the Research and Treatment of Cancer and Radiological Physics Center project. Radiation oncology, 9(123), p. 123. BioMed Central 10.1186/1748-717X-9-123
|
Text
1748-717X-9-123.pdf - Published Version Available under License Creative Commons: Attribution (CC-BY). Download (346kB) | Preview |
BACKGROUND AND PURPOSE
Intensity-modulated radiotherapy (IMRT) credentialing for a EORTC study was performed using an anthropomorphic head phantom from the Radiological Physics Center (RPC; RPC(PH)). Institutions were retrospectively requested to irradiate their institutional phantom (INST(PH)) using the same treatment plan in the framework of a Virtual Phantom Project (VPP) for IMRT credentialing.
MATERIALS AND METHODS
CT data set of the institutional phantom and measured 2D dose matrices were requested from centers and sent to a dedicated secure EORTC uploader. Data from the RPC(PH) and INST(PH) were thereafter centrally analyzed and inter-compared by the QA team using commercially available software (RIT; ver.5.2; Colorado Springs, USA).
RESULTS
Eighteen institutions participated to the VPP. The measurements of 6 (33%) institutions could not be analyzed centrally. All other centers passed both the VPP and the RPC ±7%/4 mm credentialing criteria. At the 5%/5 mm gamma criteria (90% of pixels passing), 11(92%) as compared to 12 (100%) centers pass the credentialing process with RPC(PH) and INST(PH) (p = 0.29), respectively. The corresponding pass rate for the 3%/3 mm gamma criteria (90% of pixels passing) was 2 (17%) and 9 (75%; p = 0.01), respectively.
CONCLUSIONS
IMRT dosimetry gamma evaluations in a single plane for a H&N prospective trial using the INST(PH) measurements showed agreement at the gamma index criteria of ±5%/5 mm (90% of pixels passing) for a small number of VPP measurements. Using more stringent, criteria, the RPC(PH) and INST(PH) comparison showed disagreement. More data is warranted and urgently required within the framework of prospective studies.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > Department of Haematology, Oncology, Infectious Diseases, Laboratory Medicine and Hospital Pharmacy (DOLS) > Clinic of Radiation Oncology |
UniBE Contributor: |
Weber, Damien Charles |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
1748-717X |
Publisher: |
BioMed Central |
Language: |
English |
Submitter: |
Beatrice Scheidegger |
Date Deposited: |
11 Mar 2015 15:24 |
Last Modified: |
05 Dec 2022 14:39 |
Publisher DOI: |
10.1186/1748-717X-9-123 |
PubMed ID: |
24885438 |
BORIS DOI: |
10.7892/boris.61608 |
URI: |
https://boris.unibe.ch/id/eprint/61608 |