Katsoulis, Joannis; Mericske, Regina; Rotkina, Lolita; Zbären, Christoph; Enkling, Norbert; Blatz, Markus B (2014). Precision of fit of implant-supported screw-retained 10-unit computer-aided-designed and computer-aided-manufactured frameworks made from zirconium dioxide and titanium: an in vitro study. Clinical oral implants research, 25(2), pp. 165-174. Wiley-Blackwell 10.1111/clr.12039
Text
Precision of fit.pdf - Published Version Restricted to registered users only Available under License Publisher holds Copyright. Download (725kB) |
OBJECTIVE
To analyze the precision of fit of implant-supported screw-retained computer-aided-designed and computer-aided-manufactured (CAD/CAM) zirconium dioxide (ZrO) frameworks.
MATERIALS AND METHODS
Computer-aided-designed and computer-aided-manufactured ZrO frameworks (NobelProcera) for a screw-retained 10-unit implant-supported reconstruction on six implants (FDI positions 15, 13, 11, 21, 23, 25) were fabricated using a laser (ZrO-L, N = 6) and a mechanical scanner (ZrO-M, N = 5) for digitizing the implant platform and the cuspid-supporting framework resin pattern. Laser-scanned CAD/CAM titanium (TIT-L, N = 6) and cast CoCrW-alloy frameworks (Cast, N = 5) fabricated on the same model and designed similar to the ZrO frameworks were the control. The one-screw test (implant 25 screw-retained) was applied to assess the vertical microgap between implant and framework platform with a scanning electron microscope. The mean microgap was calculated from approximal and buccal values. Statistical comparison was performed with non-parametric tests.
RESULTS
No statistically significant pairwise difference was observed between the relative effects of vertical microgap between ZrO-L (median 14 μm; 95% CI 10-26 μm), ZrO-M (18 μm; 12-27 μm) and TIT-L (15 μm; 6-18 μm), whereas the values of Cast (236 μm; 181-301 μm) were significantly higher (P < 0.001) than the three CAD/CAM groups. A monotonous trend of increasing values from implant 23 to 15 was observed in all groups (ZrO-L, ZrO-M and Cast P < 0.001, TIT-L P = 0.044).
CONCLUSIONS
Optical and tactile scanners with CAD/CAM technology allow for the fabrication of highly accurate long-span screw-retained ZrO implant-reconstructions. Titanium frameworks showed the most consistent precision. Fit of the cast alloy frameworks was clinically inacceptable.
Item Type: |
Journal Article (Original Article) |
---|---|
Division/Institute: |
04 Faculty of Medicine > School of Dental Medicine > Department of Prosthodontics [discontinued] |
UniBE Contributor: |
Katsoulis, Joannis, Mericske, Regina, Enkling, Norbert |
Subjects: |
600 Technology > 610 Medicine & health |
ISSN: |
0905-7161 |
Publisher: |
Wiley-Blackwell |
Language: |
English |
Submitter: |
Eveline Carmen Schuler |
Date Deposited: |
19 Jan 2015 16:58 |
Last Modified: |
05 Dec 2022 14:39 |
Publisher DOI: |
10.1111/clr.12039 |
PubMed ID: |
23025489 |
Uncontrolled Keywords: |
computer-aided design and computer-aided manufacturing, fixed partial denture, framework misfit, implant framework fit, implant-supported, passive fit, zirconium dioxide |
BORIS DOI: |
10.7892/boris.61744 |
URI: |
https://boris.unibe.ch/id/eprint/61744 |