In vivo compatibility of Dynesys(®) spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants

Neukamp, Michal Sarah; Röder, Christoph; Veruva, S Y; MacDonald, D W; Kurtz, S M; Steinbeck, M J (2015). In vivo compatibility of Dynesys(®) spinal implants: a case series of five retrieved periprosthetic tissue samples and corresponding implants. European spine journal, 24(5), pp. 1074-1084. Springer 10.1007/s00586-014-3705-0

[img]
Preview
Text
Neukamp_et_al_Dynesys.pdf - Published Version
Available under License Publisher holds Copyright.

Download (3MB) | Preview

PURPOSE

To determine whether particulate debris is present in periprosthetic tissue from revised Dynesys(®) devices, and if present, elicits a biological tissue reaction.

METHODS

Five Dynesys(®) dynamic stabilization systems consisting of pedicle screws (Ti alloy), polycarbonate-urethane (PCU) spacers and a polyethylene-terephthalate (PET) cord were explanted for pain and screw loosening after a mean of 2.86 years (1.9-5.3 years). Optical microscopy and scanning electron microscopy were used to evaluate wear, deformation and surface damage, and attenuated total reflectance Fourier transform infrared spectroscopy to assess surface chemical composition of the spacers. Periprosthetic tissue morphology and wear debris were determined using light microscopy, and PCU and PET wear debris by polarized light microscopy.

RESULTS

All implants had surface damage on the PCU spacers consistent with scratches and plastic deformation; 3 of 5 exhibited abrasive wear zones. In addition to fraying of the outer fibers of the PET cords in five implants, one case also evidenced cord fracture. The pedicle screws were unremarkable. Patient periprosthetic tissues around the three implants with visible PCU damage contained wear debris and a corresponding macrophage infiltration. For the patient revised for cord fracture, the tissues also contained large wear particles (>10 μm) and giant cells. Tissues from the other two patients showed comparable morphologies consisting of dense fibrous tissue with no inflammation or wear debris.

CONCLUSIONS

This is the first study to evaluate wear accumulation and local tissue responses for explanted Dynesys(®) devices. Polymer wear debris and an associated foreign-body macrophage response were observed in three of five cases.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Pre-clinic Human Medicine > Institute for Evaluative Research into Orthopaedic Surgery

UniBE Contributor:

Neukamp, Michal Sarah, Röder, Christoph

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0940-6719

Publisher:

Springer

Language:

English

Submitter:

Ives Gerber

Date Deposited:

02 Apr 2015 11:38

Last Modified:

05 Dec 2022 14:44

Publisher DOI:

10.1007/s00586-014-3705-0

PubMed ID:

25480114

BORIS DOI:

10.7892/boris.65995

URI:

https://boris.unibe.ch/id/eprint/65995

Actions (login required)

Edit item Edit item
Provide Feedback