Feasible Dose Reduction in Routine Chest Computed Tomography Maintaining Constant Image Quality Using the Last Three Scanner Generations: From Filtered Back Projection to Sinogram-affirmed Iterative Reconstruction and Impact of the Novel Fully Integrated Detector Design Minimizing Electronic Noise

Ebner, Lukas Michael; Knobloch, Felix; Huber, Adrian Thomas; Landau, Julia; Ott, Daniel; Heverhagen, Johannes; Christe, Andreas (2014). Feasible Dose Reduction in Routine Chest Computed Tomography Maintaining Constant Image Quality Using the Last Three Scanner Generations: From Filtered Back Projection to Sinogram-affirmed Iterative Reconstruction and Impact of the Novel Fully Integrated Detector Design Minimizing Electronic Noise. Journal of clinical imaging science, 4, p. 38. Medknow Publications and Media 10.4103/2156-7514.137826

[img]
Preview
Text
JClinImagingSci4138-162149_043014.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (412kB) | Preview

OBJECTIVE

The aim of the present study was to evaluate a dose reduction in contrast-enhanced chest computed tomography (CT) by comparing the three latest generations of Siemens CT scanners used in clinical practice. We analyzed the amount of radiation used with filtered back projection (FBP) and an iterative reconstruction (IR) algorithm to yield the same image quality. Furthermore, the influence on the radiation dose of the most recent integrated circuit detector (ICD; Stellar detector, Siemens Healthcare, Erlangen, Germany) was investigated.

MATERIALS AND METHODS

136 Patients were included. Scan parameters were set to a thorax routine: SOMATOM Sensation 64 (FBP), SOMATOM Definition Flash (IR), and SOMATOM Definition Edge (ICD and IR). Tube current was set constantly to the reference level of 100 mA automated tube current modulation using reference milliamperes. Care kV was used on the Flash and Edge scanner, while tube potential was individually selected between 100 and 140 kVp by the medical technologists at the SOMATOM Sensation. Quality assessment was performed on soft-tissue kernel reconstruction. Dose was represented by the dose length product.

RESULTS

Dose-length product (DLP) with FBP for the average chest CT was 308 mGy*cm ± 99.6. In contrast, the DLP for the chest CT with IR algorithm was 196.8 mGy*cm ± 68.8 (P = 0.0001). Further decline in dose can be noted with IR and the ICD: DLP: 166.4 mGy*cm ± 54.5 (P = 0.033). The dose reduction compared to FBP was 36.1% with IR and 45.6% with IR/ICD. Signal-to-noise ratio (SNR) was favorable in the aorta, bone, and soft tissue for IR/ICD in combination compared to FBP (the P values ranged from 0.003 to 0.048). Overall contrast-to-noise ratio (CNR) improved with declining DLP.

CONCLUSION

The most recent technical developments, namely IR in combination with integrated circuit detectors, can significantly lower radiation dose in chest CT examinations.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > DBMR Forschung Mu35 > Forschungsgruppe Radiologie
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > DBMR Forschung Mu35 > Forschungsgruppe Radiologie

UniBE Contributor:

Ebner, Lukas, Huber, Adrian Thomas, Landau, Julia, Ott, Daniel, Heverhagen, Johannes, Christe, Andreas

Subjects:

600 Technology > 610 Medicine & health

ISSN:

2156-7514

Publisher:

Medknow Publications and Media

Language:

English

Submitter:

Aisha Stefania Mzinga

Date Deposited:

07 Apr 2015 10:34

Last Modified:

02 Mar 2023 23:26

Publisher DOI:

10.4103/2156-7514.137826

PubMed ID:

25161807

Uncontrolled Keywords:

Dose reduction, low-dose CT, stellar detector

BORIS DOI:

10.7892/boris.66119

URI:

https://boris.unibe.ch/id/eprint/66119

Actions (login required)

Edit item Edit item
Provide Feedback