Wykrzykowska, Joanna; Serruys, Patrick; Buszman, Pawel; Linke, Axel; Ischinger, Thomas; Klauss, Volker; Eberli, Franz; Corti, Roberto; Wijns, William; Morice, Marie-Claude; Di Mario, Carlo; Van Geuns, Robert-Jan; Van Es, Gerrit-Anne; Juni, Peter; Windecker, Stephan (2011). The three year follow-up of the randomised "all-comers" trial of a biodegradable polymer biolimus-eluting stent versus permanent polymer sirolimus-eluting stent (LEADERS). EuroIntervention, 7(7), pp. 789-795. Toulouse: Europa Digital & Publishing 10.4244/EIJV7I7A125

[img]
Preview
Text
Wykrzykowska EuroIntervention 2011_p789.pdf - Published Version
Available under License Publisher holds Copyright.

Download (2MB) | Preview

Aims: The current study reports clinical outcomes at three year follow-up of the LEADERS clinical trial which was the first all-comers trial comparing a new generation biodegradable polymer biolimus drug-eluting stent (BES) with the first generation permanent polymer sirolimus-eluting stent (SES).

Methods and results: One thousand seven hundred and seven patients were randomised to unrestricted use of BES (n=857) or SES (n=850) in an all-comers population. Three year follow-up was available in 95% of the patients, 812 treated with BES and 809 treated with SES. At three years, BES remains non-inferior to SES for the primary endpoint of major adverse cardiac events (composite of cardiac death, myocardial infarction (MI), or clinically-indicated target vessel revascularisation (CI-TVR) (BES 15.7% versus SES 19%; HR 0.82 CI 0.65-1.03; p=0.09). The MACE Kaplan Meier event curves increasingly diverge with the difference in events increasing from 1.4% to 2.4% and 3.3% at 1, 2 and 3 years, respectively in favour of BES. The rate of cardiac death was non-significantly lower 4.2% versus 5.2% (HR=0.81 CI 0.52-1.26; p=0.34) and the rate of myocardial infarction was equivalent 7.2% versus 7.1% (HR 1.01 CI 0.70-1.44; p=0.97) for BES versus SES, respectively. Thus BES was non-inferior to SES in all the safety endpoints. Clinically-indicated TVR occurred in 9.4% of BES treated patients versus 11.1% of SES treated patients (HR 0.84 CI 0.62-1.13; p=0.25). Rates of definite stent thrombosis were 2.2% for BES and 2.9% for SES (HR 0.78 CI 0.43-1.43; p=0.43), with the event rate increase of 0.2% from one to three years for BES and 0.9% for SES. For patients presenting with ST-elevation myocardial infarction BES was superior to SES in reducing MACE.

Conclusions: The findings of the three year follow-up support the claim that the biodegradable polymer biolimus-eluting stent has equivalent safety and efficacy to permanent polymer sirolimus-eluting stent in an all-comers patient population. Its performance is superior in some subpopulations such as in ST-elevation MI patients and event rates for BES are overall lower than for SES with a trend toward increasing divergence of outcomes over three years.
- See more at: http://www.pcronline.com/eurointervention/42nd_issue/125/#sthash.E5HhMQ4a.dpuf

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Pre-clinic Human Medicine > Institute of Social and Preventive Medicine (ISPM)
04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Cardiology

UniBE Contributor:

Jüni, Peter, Windecker, Stephan

ISSN:

1774-024X

Publisher:

Europa Digital & Publishing

Language:

English

Submitter:

Factscience Import

Date Deposited:

04 Oct 2013 14:21

Last Modified:

05 Dec 2022 14:06

Publisher DOI:

10.4244/EIJV7I7A125

PubMed ID:

22027687

Web of Science ID:

000299609500005

BORIS DOI:

10.7892/boris.7211

URI:

https://boris.unibe.ch/id/eprint/7211 (FactScience: 212396)

Actions (login required)

Edit item Edit item
Provide Feedback