Feasibility of cardiopulmonary exercise testing and training using a robotics-assisted tilt table in dependent-ambulatory stroke patients

Saengsuwan, Jittima; Huber, Céline; Schreiber, Jonathan; Schuster-Amft, Corina; Nef, Tobias; Kenneth, J. Hunt (2015). Feasibility of cardiopulmonary exercise testing and training using a robotics-assisted tilt table in dependent-ambulatory stroke patients. Journal of NeuroEngineering and Rehabilitation, 12(88), p. 88. BioMed Central 10.1186/s12984-015-0078-5

[img]
Preview
Text
s12984-015-0078-5.pdf - Published Version
Available under License Creative Commons: Attribution (CC-BY).

Download (925kB) | Preview

BACKGROUND: We evaluated the feasibility of an augmented robotics-assisted tilt table (RATT) for incremental cardiopulmonary exercise testing (CPET) and exercise training in dependent-ambulatory stroke patients.
METHODS: Stroke patients (Functional Ambulation Category ≤ 3) underwent familiarization, an incremental exercise test (IET) and a constant load test (CLT) on separate days. A RATT equipped with force sensors in the thigh cuffs, a work rate estimation algorithm and real-time visual feedback to guide the exercise work rate was used. Feasibility assessment considered technical feasibility, patient tolerability, and cardiopulmonary responsiveness.
RESULTS: Eight patients (4 female) aged 58.3 ± 9.2 years (mean ± SD) were recruited and all completed the study. For IETs, peak oxygen uptake (V'O2peak), peak heart rate (HRpeak) and peak work rate (WRpeak) were 11.9 ± 4.0 ml/kg/min (45 % of predicted V'O2max), 117 ± 32 beats/min (72 % of predicted HRmax) and 22.5 ± 13.0 W, respectively. Peak ratings of perceived exertion (RPE) were on the range "hard" to "very hard". All 8 patients reached their limit of functional capacity in terms of either their cardiopulmonary or neuromuscular performance. A ventilatory threshold (VT) was identified in 7 patients and a respiratory compensation point (RCP) in 6 patients: mean V'O2 at VT and RCP was 8.9 and 10.7 ml/kg/min, respectively, which represent 75 % (VT) and 85 % (RCP) of mean V'O2peak. Incremental CPET provided sufficient information to satisfy the responsiveness criteria and identification of key outcomes in all 8 patients. For CLTs, mean steady-state V'O2 was 6.9 ml/kg/min (49 % of V'O2 reserve), mean HR was 90 beats/min (56 % of HRmax), RPEs were > 2, and all patients maintained the active work rate for 10 min: these values meet recommended intensity levels for bouts of training.
CONCLUSIONS: The augmented RATT is deemed feasible for incremental cardiopulmonary exercise testing and exercise training in dependent-ambulatory stroke patients: the approach was found to be technically implementable, acceptable to the patients, and it showed substantial cardiopulmonary responsiveness. This work has clinical implications for patients with severe disability who otherwise are not able to be tested.

Item Type:

Journal Article (Original Article)

Division/Institute:

10 Strategic Research Centers > ARTORG Center for Biomedical Engineering Research > ARTORG Center - Gerontechnology and Rehabilitation
10 Strategic Research Centers > ARTORG Center for Biomedical Engineering Research

UniBE Contributor:

Saengsuwan, Jittima, Nef, Tobias

Subjects:

500 Science > 570 Life sciences; biology
600 Technology > 610 Medicine & health
600 Technology > 620 Engineering

ISSN:

1743-0003

Publisher:

BioMed Central

Language:

English

Submitter:

Vanessa Vallejo

Date Deposited:

09 Oct 2015 13:28

Last Modified:

05 Dec 2022 14:49

Publisher DOI:

10.1186/s12984-015-0078-5

PubMed ID:

26410821

BORIS DOI:

10.7892/boris.72238

URI:

https://boris.unibe.ch/id/eprint/72238

Actions (login required)

Edit item Edit item
Provide Feedback