European multicentre experience with the sutureless Perceval valve: clinical and haemodynamic outcomes up to 5 years in over 700 patients†.

Shrestha, Malakh; Fischlein, Theodore; Meuris, Bart; Flameng, Willem; Carrel, Thierry; Madonna, Francesco; Misfeld, Martin; Folliguet, Thierry; Haverich, Axel; Laborde, Francois (2016). European multicentre experience with the sutureless Perceval valve: clinical and haemodynamic outcomes up to 5 years in over 700 patients†. European journal of cardio-thoracic surgery, 49(1), pp. 234-241. Oxford University Press 10.1093/ejcts/ezv040

[img]
Preview
Text
234.full.pdf - Published Version
Available under License Publisher holds Copyright.

Download (193kB) | Preview
[img]
Preview
Text
Paper.without.figures. Perceval EACTS 2014_unmarked revision final_24.09.2014_clean.pdf - Accepted Version
Available under License Publisher holds Copyright.

Download (172kB) | Preview

OBJECTIVES

This report summarizes the 5-year clinical and haemodynamic data from three prospective, European multicentre trials with the Perceval sutureless aortic valve.

METHODS

From April 2007 to August 2012, 731 consecutive patients (mean age: 78.5 years; 68.1% females; mean logistic EuroSCORE 10.9%) underwent AVR with the Perceval valve in 25 European centres. Isolated AVR was performed in 498 (68.1%) patients. A minimally invasive approach was performed in 189 (25.9%) cases. The cumulative follow-up was 729 patients-years.

RESULTS

In isolated AVR, mean cross-clamp and cardiopulmonary bypass times were 30.8 and 50.8 min in full sternotomy, and 37.6 and 64.4 min in the minimally invasive approach, respectively. Early cardiac-related deaths occurred in 1.9%. Overall survival rates at 1 and 5 years were 92.1 and 74.7%, respectively. Major paravalvular leak occurred in 1.4% and 1% at early and late follow-up, respectively. Significant improvement in clinical status was observed postoperatively in the majority of patients. Mean and peak gradients decreased from 42.9 and 74.0 mmHg preoperatively, to 7.8 and 16 mmHg at the 3-year follow-up. LV mass decreased from 254.5 to 177.4 g at 3 years.

CONCLUSIONS

This European multicentre experience, with the largest cohort of patients with sutureless valves to date, shows excellent clinical and haemodynamic results that remain stable even up to the 5-year follow-up. Even in this elderly patient cohort with 40% octogenarians, both early and late mortality rates were very low. There were no valve migrations, structural valve degeneration or valve thrombosis in the follow-up. The sutureless technique is a promising alternative to biological aortic valve replacement.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Cardiovascular Disorders (DHGE) > Clinic of Heart Surgery

UniBE Contributor:

Carrel, Thierry

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1010-7940

Publisher:

Oxford University Press

Language:

English

Submitter:

Daniela Huber

Date Deposited:

17 Feb 2016 10:24

Last Modified:

27 Feb 2024 14:28

Publisher DOI:

10.1093/ejcts/ezv040

PubMed ID:

25750010

Uncontrolled Keywords:

Aortic valve replacement; Prospective study; Sutureless heart valve prosthesis

BORIS DOI:

10.7892/boris.75623

URI:

https://boris.unibe.ch/id/eprint/75623

Actions (login required)

Edit item Edit item
Provide Feedback