Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA.

Filli, Lukas; Piccirelli, Marco; Kenkel, David; Boss, Andreas; Manoliu, Andrei; Andreisek, Gustav; Bhat, Himanshu; Runge, Val Murray; Guggenberger, Roman (2016). Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA. European radiology, 26(6), pp. 1921-1928. Springer 10.1007/s00330-015-3985-8

[img]
Preview
Text
art%3A10.1007%2Fs00330-015-3985-8.pdf - Published Version
Available under License Publisher holds Copyright.

Download (1MB) | Preview

PURPOSE

To investigate the feasibility of MR diffusion tensor imaging (DTI) of the median nerve using simultaneous multi-slice echo planar imaging (EPI) with blipped CAIPIRINHA.

MATERIALS AND METHODS

After federal ethics board approval, MR imaging of the median nerves of eight healthy volunteers (mean age, 29.4 years; range, 25-32) was performed at 3 T using a 16-channel hand/wrist coil. An EPI sequence (b-value, 1,000 s/mm(2); 20 gradient directions) was acquired without acceleration as well as with twofold and threefold slice acceleration. Fractional anisotropy (FA), mean diffusivity (MD) and quality of nerve tractography (number of tracks, average track length, track homogeneity, anatomical accuracy) were compared between the acquisitions using multivariate ANOVA and the Kruskal-Wallis test.

RESULTS

Acquisition time was 6:08 min for standard DTI, 3:38 min for twofold and 2:31 min for threefold acceleration. No differences were found regarding FA (standard DTI: 0.620 ± 0.058; twofold acceleration: 0.642 ± 0.058; threefold acceleration: 0.644 ± 0.061; p ≥ 0.217) and MD (standard DTI: 1.076 ± 0.080 mm(2)/s; twofold acceleration: 1.016 ± 0.123 mm(2)/s; threefold acceleration: 0.979 ± 0.153 mm(2)/s; p ≥ 0.074). Twofold acceleration yielded similar tractography quality compared to standard DTI (p > 0.05). With threefold acceleration, however, average track length and track homogeneity decreased (p = 0.004-0.021).

CONCLUSION

Accelerated DTI of the median nerve is feasible. Twofold acceleration yields similar results to standard DTI.

KEY POINTS

• Standard DTI of the median nerve is limited by its long acquisition time. • Simultaneous multi-slice acquisition is a new technique for accelerated DTI. • Accelerated DTI of the median nerve yields similar results to standard DTI.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic, Interventional and Paediatric Radiology

UniBE Contributor:

Runge, Val Murray

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0938-7994

Publisher:

Springer

Language:

English

Submitter:

Aisha Stefania Mzinga

Date Deposited:

13 Apr 2016 16:26

Last Modified:

05 Dec 2022 14:52

Publisher DOI:

10.1007/s00330-015-3985-8

PubMed ID:

26373759

Uncontrolled Keywords:

Diffusion tensor imaging; Diffusion tractography; Echo-planar imaging; Median nerve; Simultaneous multi-slice

BORIS DOI:

10.7892/boris.77239

URI:

https://boris.unibe.ch/id/eprint/77239