Comparison of Routine Brain Imaging at 3 T and 7 T.

Springer, Elisabeth; Dymerska, Barbara; Cardoso, Pedro Lima; Robinson, Simon Daniel; Weisstanner, Christian; Wiest, Roland; Schmitt, Benjamin; Trattnig, Siegfried (2016). Comparison of Routine Brain Imaging at 3 T and 7 T. Investigative radiology, 51(8), p. 1. Lippincott Williams & Wilkins 10.1097/RLI.0000000000000256

[img]
Preview
Text
00004424-900000000-99227.pdf - Published Version
Available under License Publisher holds Copyright.

Download (1MB) | Preview

OBJECTIVE

The aim of this study was to compare quantitative and semiquantitative parameters (signal-to-noise ratio [SNR], contrast-to-noise ratio [CNR], image quality, diagnostic confidence) from a standard brain magnetic resonance imaging examination encompassing common neurological disorders such as demyelinating disease, gliomas, cerebrovascular disease, and epilepsy, with comparable sequence protocols and acquisition times at 3 T and at 7 T.

MATERIALS AND METHODS

Ten healthy volunteers and 4 subgroups of 40 patients in total underwent comparable magnetic resonance protocols with standard diffusion-weighted imaging, 2D and 3D turbo spin echo, 2D and 3D gradient echo and susceptibility-weighted imaging of the brain (10 sequences) at 3 T and 7 T. The subgroups comprised patients with either lesional (n = 5) or nonlesional (n = 4) epilepsy, intracerebral tumors (n = 11), demyelinating disease (n = 11) (relapsing-remitting multiple sclerosis [MS, n = 9], secondary progressive MS [n = 1], demyelinating disease not further specified [n = 1]), or chronic cerebrovascular disorders [n = 9]). For quantitative analysis, SNR and CNR were determined. For a semiquantitative assessment of the diagnostic confidence, a 10-point scale diagnostic confidence score (DCS) was applied. Two experienced radiologists with additional qualification in neuroradiology independently assessed, blinded to the field strength, 3 pathology-specific imaging criteria in each of the 4 disease groups and rated their diagnostic confidence. The overall image quality was semiquantitatively assessed using a 4-point scale taking into account whether diagnostic decision making was hampered by artifacts or not.

RESULTS

Without correction for spatial resolution, SNR was higher at 3 T except in the T2 SPACE 3D, DWI single shot, and DIR SPACE 3D sequences. The SNR corrected by the ratio of 3 T/7 T voxel sizes was higher at 7 T than at 3 T in 10 of 11 sequences (all except for T1 MP2RAGE 3D).In CNR, there was a wide variation between sequences and patient cohorts, but average CNR values were broadly similar at 3 T and 7 T.DCS values for all 4 pathologic entities were higher at 7 T than at 3 T. The DCS was significantly higher at 7 T for diagnosis and exclusion of cortical lesions in vascular disease. A tendency to higher DCS at 7 T for cortical lesions in MS was observed, and for the depiction of a central vein and iron deposits within MS lesions. Despite motion artifacts, DCS values were higher at 7 T for the diagnosis and exclusion of hippocampal sclerosis in mesial temporal lobe epilepsy (improved detection of the hippocampal subunits). Interrater agreement was 69.7% at 3 T and 93.3% at 7 T. There was no significant difference in the overall image quality score between 3 T and 7 T taking into account whether diagnostic decision making was hampered by artifacts or not.

CONCLUSIONS

Ultra-high-field magnetic resonance imaging at 7 T compared with 3 T yielded an improved diagnostic confidence in the most frequently encountered neurologic disorders. Higher spatial resolution and contrast were identified as the main contributory factors.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > Department of Radiology, Neuroradiology and Nuclear Medicine (DRNN) > Institute of Diagnostic and Interventional Neuroradiology

UniBE Contributor:

Springer, Elisabeth, Weisstanner, Christian, Wiest, Roland Gerhard Rudi

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0020-9996

Publisher:

Lippincott Williams & Wilkins

Language:

English

Submitter:

Martin Zbinden

Date Deposited:

09 Jun 2016 08:21

Last Modified:

02 Mar 2023 23:27

Publisher DOI:

10.1097/RLI.0000000000000256

PubMed ID:

26863580

BORIS DOI:

10.7892/boris.81240

URI:

https://boris.unibe.ch/id/eprint/81240

Actions (login required)

Edit item Edit item
Provide Feedback