Analytic validation of commercially available immunoassays for the measurement of serum cobalamin and folate concentrations in pigs

Grützner, Niels; Knabe, Darrel; Lawhorn, Bruce D; Dominguez, Brandon; Kauffold, Johannes; Suchodolski, Jan S; Steiner, Jörg M (2016). Analytic validation of commercially available immunoassays for the measurement of serum cobalamin and folate concentrations in pigs. Veterinary clinical pathology, 45(2), pp. 311-319. Wiley-Blackwell 10.1111/vcp.12361

[img] Text
vcp12361.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (227kB)

BACKGROUND

Cobalamin (vitamin B12 ) and folate (vitamin B9 ) are important for the amino acid metabolism and nucleic acid synthesis. Immunoassays for the measurement of both vitamins in serum are routinely used in people, cats, and dogs, serving as indicators for clinical disorders including cobalamin and/or folate deficiency, small intestinal dysbiosis, or inadequate dietary supply of these vitamins. The analysis of these analytes may also be clinically useful in pigs.

OBJECTIVE

The purpose of this study was to analytically validate immunoassays for the measurement of cobalamin and folate concentrations in porcine serum, and to determine serum cobalamin and folate concentrations in healthy newborn pigs pre- and postweaning.

METHODS

Assay validation for both vitamins included the determination of linearity, accuracy, and intra- and inter-assay variability using serum samples from 10 pigs. Also, serum cobalamin and folate concentrations were compared in piglets between pre- and postweaning.

RESULTS

For both vitamins, observed-to-expected ratios for linearity and accuracy were 93.2 ± 14.3% and 100.3 ± 8.1% (mean ± standard deviation), respectively. Intra- and inter-assay coefficients of variation for serum were ≤ 8.7% and ≤12.5%, respectively. Significantly higher serum cobalamin and lower folate concentrations were observed in piglets at the time of weaning than at postweaning (P < .0061; P < .0001, respectively).

CONCLUSIONS

Both immunoassays are linear, accurate, precise, and reproducible for measurement of porcine serum cobalamin and folate concentrations. Piglets differing in age by only 12 days had significantly different serum cobalamin and folate concentrations. The implications of differing serum cobalamin and folate concentrations in pigs at different stages of life should be further investigated.

Item Type:

Journal Article (Original Article)

Division/Institute:

05 Veterinary Medicine > Research Foci > Veterinary Public Health / Herd Health Management
05 Veterinary Medicine > Department of Clinical Veterinary Medicine (DKV) > Swine Clinic

UniBE Contributor:

Grützner, Niels

Subjects:

600 Technology > 610 Medicine & health

ISSN:

0275-6382

Publisher:

Wiley-Blackwell

Language:

English

Submitter:

Jeannette Glauser

Date Deposited:

29 Aug 2016 08:29

Last Modified:

05 Dec 2022 14:57

Publisher DOI:

10.1111/vcp.12361

PubMed ID:

27153102

Uncontrolled Keywords:

Plasma; sera; swine; vitamin B12; vitamin B9

BORIS DOI:

10.7892/boris.85833

URI:

https://boris.unibe.ch/id/eprint/85833

Actions (login required)

Edit item Edit item
Provide Feedback