Oxygen-transfer performance of a newly designed, very low-volume membrane oxygenator

Burn, Felice Antonio (2015). Oxygen-transfer performance of a newly designed, very low-volume membrane oxygenator. (Dissertation, Insel-Gruppe, Medizin)

Full text not available from this repository. (Request a copy)

Gebiet: Chirurgie Biomedizintechnik Biophysik Transplantationsmedizin Kardiologie Abstract: OBJECTIVES: – Oxygenation of blood and other physiological solutions are routinely required in fundamental research for both in vitro and in vivo experimentation. However, very few oxygenators with suitable priming volumes (<2-3 ml) are available for surgery in small animals. We have designed a new, miniaturized membrane oxygenator and investigated the oxygen-transfer performance using both buffer and blood perfusates. – – METHODS: – The mini-oxygenator was designed with a central perforated core-tube surrounded by parallel-oriented microporous polypropylene hollow fibres, placed inside a hollow shell with a lateral-luer outlet, and sealed at both extremities. With this design, perfusate is delivered via the core-tube to the centre of the mini-oxygenator, and exits via the luer port. A series of mini-oxygenators were constructed and tested in an in vitro perfusion circuit by monitoring oxygen transfer using modified Krebs-Henseleit buffer or whole porcine blood. Effects of perfusion pressure and temperature over flows of 5-60 ml × min(-1) were assessed. – – RESULTS: – Twelve mini-oxygenators with a mean priming volume of 1.5 ± 0.3 ml were evaluated. With buffer, oxygen transfer reached a maximum of 14.8 ± 1.0 ml O2 × l(-1) (pO2: 450 ± 32 mmHg) at perfusate flow rates of 5 ml × min(-1) and decreased with an increase in perfusate flow to 7.8 ± 0.7 ml ml O2 × l(-1) (pO2: 219 ± 24 mmHg) at 60 ml × min(-1). Similarly, with blood perfusate, oxygen transfer also decreased as perfusate flow increased, ranging from 33 ± 5 ml O2 × l(-1) at 5 ml × min(-1) to 11 ± 2 ml O2 × l(-1) at 60 ml × min(-1). Furthermore, oxygen transfer capacity remained stable with blood perfusion over a period of at least 2 h. – – CONCLUSIONS: – We have developed a new miniaturized membrane oxygenator with an ultra-low priming volume (<2 ml) and adequate oxygenation performance. This oxygenator may be of use in overcoming current limitations in equipment size for effective oxygenation in low-volume perfusion circuits, such as small animal extracorporeal circulation and ex vivo organ perfusion. – – © The Author 2015. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Item Type:

Thesis (Dissertation)

Division/Institute:

04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > DBMR Forschung Mu35 > Forschungsgruppe Herz- und Gefässchirurgie
04 Faculty of Medicine > Pre-clinic Human Medicine > BioMedical Research (DBMR) > DBMR Forschung Mu35 > Forschungsgruppe Herz- und Gefässchirurgie

04 Faculty of Medicine > Department of Head Organs and Neurology (DKNS) > Clinic of Neurosurgery

UniBE Contributor:

Burn, Felice Antonio, Henning Longnus, Sarah

Subjects:

600 Technology > 610 Medicine & health

Language:

English

Submitter:

Daniela Huber

Date Deposited:

18 Feb 2016 16:56

Last Modified:

05 Dec 2022 14:51

URI:

https://boris.unibe.ch/id/eprint/75864

Actions (login required)

Edit item Edit item
Provide Feedback