Implants with Original and Non-Original Abutment Connections

Gigandet, Michel; Gianni, Bigolin; Francisco, Faoro; Bürgin, Walter Bruno; Brägger, Urs (2014). Implants with Original and Non-Original Abutment Connections. Clinical implant dentistry and related research, 16(2), pp. 303-311. Oxford: Blackwell 10.1111/j.1708-8208.2012.00479.x

[img] Text
cid479.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (962kB) | Request a copy

AIM: To test in vitro the mechanical resistance, rotational misfit and failure mode of three original implant-abutment connections and to compare them to two connections between non-original abutments connected to one of the original implants. MATERIAL AND METHODS: Three different implants with small diameters (3.3 mm for Straumann Roxolid, 3.5 mm for Nobel Biocare Replace and Astra Tech Osseospeed TX) were connected with individualized titanium abutments. Twelve implants from each system were connected to their original abutments (Straumann CARES, Nobel Biocare Procera, Astra Tech Atlantis). Twenty-four Roxolid implants were connected with non-original abutments using CAD/CAM procedures from the other two manufacturers (12 Nobel Biocare Procera and 12 Astra Tech Atlantis). For the critical bending test, a Zwick/Roell 1475 machine and the Xpert Zwick/Roell software were used. RESULTS: The rotational misfit varied when comparing the different interfaces. The use of non-original grade V titanium abutments on Roxolid implants increased the force needed for deformation. The fracture mode was different with one of the original connections. CONCLUSIONS: Non-original abutments differ in design of the connecting surfaces and material and demonstrate higher rotational misfit. These differences may result in unexpected failure modes.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology
04 Faculty of Medicine > School of Dental Medicine > Division of Fixed Prosthodontics [discontinued]
04 Faculty of Medicine > School of Dental Medicine > Research
04 Faculty of Medicine > School of Dental Medicine > Department of Preventive, Restorative and Pediatric Dentistry

UniBE Contributor:

Bürgin-Mumenthaler, Walter Bruno, Brägger, Urs

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1523-0899

Publisher:

Blackwell

Language:

English

Submitter:

Eveline Carmen Schuler

Date Deposited:

04 Oct 2013 14:30

Last Modified:

02 Mar 2023 23:21

Publisher DOI:

10.1111/j.1708-8208.2012.00479.x

PubMed ID:

22804848

Uncontrolled Keywords:

abutment connection, abutments, CAD/CAM, dental implants, ISO 14801, non-original, original, titanium grade IV, titanium grade V

BORIS DOI:

10.7892/boris.11336

URI:

https://boris.unibe.ch/id/eprint/11336 (FactScience: 217429)

Actions (login required)

Edit item Edit item
Provide Feedback