Clinical Performance of One-Piece, Screw-Retained Implant Crowns Based on Hand-Veneered CAD/CAM Zirconia Abutments After a Mean Follow-up Period of 2.3 Years.

Schnider, Nicole; Forrer, Fiona; Brägger, Urs; Hicklin, Stefan Paul (2018). Clinical Performance of One-Piece, Screw-Retained Implant Crowns Based on Hand-Veneered CAD/CAM Zirconia Abutments After a Mean Follow-up Period of 2.3 Years. The international journal of oral & maxillofacial implants, 33(1), pp. 188-196. Quintessence Publishing 10.11607/jomi.5929

[img] Text
jomi_2018_01_s0188.pdf - Published Version
Restricted to registered users only
Available under License Publisher holds Copyright.

Download (210kB)

PURPOSE

The aim of this study was to evaluate the clinical performance of one-piece, screw-retained implant crowns based on hand-veneered computer-aided design/computer-aided manufacture (CAD/CAM) zirconium dioxide abutments with a crossfit connection at least 1 year after insertion of the crown.

MATERIALS AND METHODS

Consecutive patients who had received at least one Straumann bone level implant and one-piece, screw-retained implant crowns fabricated with CARES zirconium dioxide abutments were reexamined. Patient satisfaction, occlusal and peri-implant parameters, mechanical and biologic complications, radiologic parameters, and esthetics were recorded.

RESULTS

A total of 50 implant crowns in the anterior and premolar region were examined in 41 patients. The follow-up period of the definitive reconstructions ranged from 1.1 to 3.8 years. No technical and no biologic complications had occurred. At the reexamination, 100% of the implants and reconstructions were in situ. Radiographic evaluation revealed a mean distance from the implant shoulder to the first visible bone-to-implant contact of 0.06 mm at the follow-up examination.

CONCLUSION

Screw-retained crowns based on veneered CAD/CAM zirconium dioxide abutments with a crossfit connection seem to be a promising way to replace missing teeth in the anterior and premolar region. In the short term, neither failures of components nor complications were noted, and the clinical and radiographic data revealed stable hard and soft tissue conditions.

Item Type:

Journal Article (Original Article)

Division/Institute:

04 Faculty of Medicine > School of Dental Medicine > Department of Reconstructive Dentistry and Gerodontology

UniBE Contributor:

Schnider, Nicole, Forrer, Fiona, Brägger, Urs

Subjects:

600 Technology > 610 Medicine & health

ISSN:

1942-4434

Publisher:

Quintessence Publishing

Language:

English

Submitter:

Vanda Kummer

Date Deposited:

22 Feb 2019 09:18

Last Modified:

05 Dec 2022 15:24

Publisher DOI:

10.11607/jomi.5929

PubMed ID:

29340353

BORIS DOI:

10.7892/boris.123967

URI:

https://boris.unibe.ch/id/eprint/123967

Actions (login required)

Edit item Edit item
Provide Feedback